NI 446X Calibration Procedure
6.
Configure the timing properties of the acquisition using the following
function.
7.
Start the task using the following function.
NI-DAQmx Function Call
LabVIEW Block Diagram
Call
DAQmxCfgSampClkTiming
with the
following parameters:
taskHandle
:
myTaskHandle
source
:
NULL
rate
:
204800.0
activeEdge
:
DAQmx_Val_Rising
sampleMode
:
DAQmx_Val_FiniteSamps
sampsPerChanToAcquire
:
102400
NI-DAQmx Function Call
LabVIEW Block Diagram
Call
DAQmxStartTask
with the following
parameters:
taskHandle
:
myTaskHandle