National Instruments Corporation
57
NI 4065 Calibration Procedure
5.
Repeat step 4 for each of the remaining iterations listed in Table 15.
6.
Reset the calibrator.
Adjusting DC Voltage
To adjust DC voltage for the NI 4065, complete the following steps:
1.
Call the niDMM Reset VI.
Table 15.
Linearization Settings
Iteration
Calibrator Output
niDMM Cal Adjust Linearization Parameters
Range (V)
Expected Value
Amplitude (V)
1
–1.0
1
–
1.0
2
–0.8
1
–
0.8
3
–0.6
1
–
0.6
4
–0.4
1
–
0.4
5
–0.2
1
–
0.2
6
0.0
1
0.0
7
0.2
1
0.2
8
0.4
1
0.4
9
0.6
1
0.6
10
0.8
1
0.8
11
1.0
1
1.0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_reset
with the following
parameter:
Instrument_Handle
:
Cal Session