4. Equipment overview
The UV unit contains of:
CleanMove 8 for surface and air disinfection.
4.1 Environment for the equipment
4.1.1 Surface
The installation must be placed on a flat surface without inclination and fastened according
to the instructions.
4.1.2. Placement
The equipment may only be used in places where the surface is level and can be securely fastened.
4.1.3 Temperature and humidity
The equipment can be used in temperatures between 10-40°C and max. 60% humidity as standard.
4.1.4. Storage and environment
The equipment is not designed for outdoor use or storage.
If the equipment comes into contact with saline or acidic moisture or liquid, this should be removed from the
system as soon as possible to prevent corrosion.
4.1.5. Requirement before use
The equipment should be used as a disinfection solution with the purchaser:
•
The equipment must not be used until the system has been completely erected and installed
in accordance with the regulations.
•
The equipment may only be operated by personnel who have been properly instructed in its
use.
•
Repairs, service, and maintenance may only be carried out by qualified persons with proper
training.
•
If constructive changes are made to the equipment, the CE marking will lapse.
4.1.6 Disposal of parts
Consumables must be disposed of in accordance with current legislation.
•
UVC-
lamps contain mercury and must be disposed of in accordance with current regulations
for lamps containing mercury.
4.2. Mounting instructions
Installation of UV systems must be done according to specifications or further agreement.
The UVC light is outside the visible spectrum and therefore bulbs should be replaced according to the
following instructions.
•
Installations with light 24/7 in normal temperature
ranges
Change after 14 months
of continuous operation
•
Installations with light 24/7 in temperatures
>40C eller <10C
Change after 12 months
of continuous operation
•
Installations with frequent switch on and off
Max. 3000 on/off
•
Installations with other type of use
Max. 2 years from delivery date
Avoid short-term (<5 min.) Switching on / off of bulbs. It has a strong degenerative effect on the life of the
bulbs and is not covered by the warranty obligation.