2
The Advanced “Airway Larry” Manikin is a
Complete Resuscitation System consisting
of modular components which allow you
to create a manikin to suit your changing
needs. The components may be pur-
chased as a complete package or sepa-
rately to update your existing manikin.
Update packages are compatible with all
versions of both
CPARLENE
®
and non-
modular Resusci
®
Anne™*.
This manual will guide you in setting
up, using, and maintaining each of the
available components. Each section also
includes a list of replacement parts, sup-
plies, and auxiliary equipment.
By reading and following all instructions
carefully and completely, you can be
sure your
Advanced “Airway
Larry” Manikin will provide years of valu-
able service.
Cleaning:
Normal surface soil can be removed
from the trainer with mild soapy water.
Stubborn stains may be removed with
REN Cleaner (W09919U). Simply apply
the REN Cleaner to the soiled area and
wipe clean with a soft cloth.
Note:
Avoid using cleaner around the
mouth area if students will be apply-
ing direct mouth-to-mouth resuscitation
techniques, as the cleaner may be toxic if
ingested. NEVER place the trainer on any
kind of printed paper or plastic. These
materials, as well as ballpoint pens, will
transfer indelible stains. Do not use any
cosmetics.
List of Components:
• Advanced “Airway Larry”
Torso Manikin
• Defibrillation Chest Skin
• 8-oz. Pump Spray Lubricant
• 20 cc syringe (2)
*Resusci
®
Anne™ is a trademark of Laerdal Medical Corporation.
NASCO Defibrillation
Chest Skin
About the Simulator…
The
CR
i
S
is
™
System Defibrillation Chest
Skin has been designed to absorb a
maximum of 360 joules of energy*.
Although capable of absorbing 360
joules, we do recommend that the small-
est energy level possible be used while
training with the skin.
The
CR
i
S
is
™
System Defibrillation Chest
Skin will enable you to practice defibril-
lation using manual, semi-automatic,
and automatic external defibrillators
(AEDs). When using any one of these in
training, always follow the recommended
operating procedures for that particular
defibrillator.
The “Airway Larry” Manikin comes with
the Defibrillation Chest Skin in place and
ready for use. Connect the defibrillation
chest skin to your arrhythmia or patient
simulator using the four-lead snap cable.
Connecting Your Arrhythmia or
Patient Simulator:
Nasco has designed the
CR
i
S
is
™
System
to be compatible with a variety of patient
simulators. This is possible via the stan-
dard four-lead snap cable (Figure 1). If
your patient simulator has only two out-
put posts, the red and black leads must
be connected to the patient simulator.
You may want to connect the manikin
to the simulator that came with your
defibrillator. If your patient simulator
doesn’t have the standard snap connec-
tors, it will be necessary to purchase the
corresponding adapters from Nasco. To
order, please see “Supplies/Replacement
Parts” on page 8.
Figure 1
7
of 360 joules may cause irreparable
damage to the chest skin, circuitry,
and patient simulator being used
— thus voiding Nasco’s warranty and
endangering your equipment. Nasco
assumes no liability for damage or
injury that may be caused by the use
and/or misuse of this equipment. All
normal safety precautions for defibril-
lation training should be followed and
energy levels should be minimized.
Nasco did not design nor intend this
Defibrillation Chest Skin to be used as
anything other than a training appa-
ratus for defibrillation.
Note:
360 joules is the maximum
energy level that Nasco recommends
administering to the Defibrillation
Chest Skin. Energy levels in excess
NP 154-06 LF03960.indd 4