17 Specifications
Narda
SRM-3006
203
17.7
Single axis H-field antenna 3551/02
17.7.1
Characteristics
1 Typical measurement dynamic range for 10 dB signal to noise radio (RBW = 1 kHz); for
frequencies > 10 MHz
2 Typical values
17.7.2
Uncertainty
1 Typical value k = 2 (K= extrapolation or correction factor for determining the assessment value);
+15 °C to +30 °C
17.7.3
General specifications
Frequency range
9 kHz – 300 MHz
The correction factors determined individually
during calibration are stored in an EEPROM and
are applied automatically when used in conjunction
with the SRM-3006 basic unit.
Antenna type
H-field
Sensor type
Single axis active magnetic loop
Dynamic range
1)
0.4 µA/m – 71 mA/m
CW damage level
> 2.65 A/m above 1 MHz
Intrinsic noise display in
conjunction with the SRM-3006
basic unit
2)
0.12 µA/m for each frequency > 10 MHz with 1 kHz
resolution bandwidth RBW
Measurement range limit
(for single CW signal)
2)
100 mA/m
RF connector
N connector, 50
Ω
Extended measurement
uncertainty
1)
(in conjunction with SRM-3006
basic unit and 1.5 m RF cable)
Frequency range
Single axis
measurement
0.1 – 300 MHz
2.0 dB
Calibration uncertainty
< 1.2 dB
Operating temperature range
-10 °C to +50 °C (same as SRM-3006 basic unit)
Air humidity
< 29 g/m
3
(< 93 % at +30 °C)
Summary of Contents for SRM-3006
Page 1: ...SRM 3006 Selective Radiation Meter Operating Manual ...
Page 12: ...Contents X SRM 3006 Narda ...
Page 15: ...1 Introduction Narda SRM 3006 3 Figure 1 SRM 3006 in on site use ...
Page 24: ...1 Introduction 12 SRM 3006 Narda ...
Page 50: ...3 Connecting Up and Starting to Use the Device 38 SRM 3006 Narda ...
Page 64: ...4 Operation and Basic Settings 52 SRM 3006 Narda ...
Page 128: ...8 Spectrum Analysis Mode 116 SRM 3006 Narda ...
Page 146: ...10 Scope Mode 134 SRM 3006 Narda ...
Page 168: ...12 Spatial Averaging 156 SRM 3006 Narda ...
Page 178: ...13 Measurement Data Memory 166 SRM 3006 Narda ...
Page 182: ...14 Maintenance and Repairs 170 SRM 3006 Narda ...
Page 186: ...15 PC Software 174 SRM 3006 Narda ...
Page 192: ...16 Remote Control 180 SRM 3006 Narda ...
Page 217: ...17 Specifications Narda SRM 3006 205 17 8 Declaration of Conformity ...
Page 218: ...17 Specifications 206 SRM 3006 Narda ...
Page 224: ...18 Annex A 212 SRM 3006 Narda ...
Page 230: ...19 Ordering Information 218 SRM 3006 Narda ...
Page 238: ...Index 226 SRM 3006 Narda ...
Page 239: ......