4.11 Remote control
Narda
NBM-520
31
To set the alarm limit
For normal (flat) probes:
⇒
Select the desired limit value for
LIMIT FLAT
using the
PC software.
ª
The symbol
is shown in the display when the
alarm function is activated (see
Table 3
Setting ranges of the limit values (in 1 dB steps)
For shaped probes:
⇒
Select the desired limit value (percentage of the stan-
dard value) for
LIMIT SHAP
using the PC software.
Setting range: 0.1% to 10,000% (in 1 dB steps)
ª
The symbol
is shown in the display when the
alarm function is activated (see
To deactivate the alarm function
⇒
Set the parameter
ALARM
to
OFF
using the PC soft-
ware.
4.11 Remote control
The NBM-520 can be remote controlled using the NBM-TS
software or from an NBM-550 measuring set. This allows
you to use the NBM-520 in places that are difficult to reach
and read the results without needing to see the actual dis-
play on the NBM-520. While the NBM-520 is remote con-
trolled,
REMOTE
appears in the display instead of field type
(E-Field or H-Field).
Measuring range Minimum
Maximum
V/m 100 mV/m
100 kV/m
A/m 250
μ
A/m
250 A/m
W/m
2
25
μ
W/m
2
25 MW//m
2
mW/cm
2
2.5 nW/cm
2
2.5 kW/cm
2
Summary of Contents for nbm-520
Page 1: ...NBM 520 Narda Broadband Field Meter Operating Manual...
Page 6: ...Contents iv NBM 520 Narda...
Page 12: ...1 Useful information 6 NBM 520 Narda...
Page 21: ...3 2 Instrument overview Narda NBM 520 15 3 2 Instrument overview 7 1 2 3 4 9 10 6a 6b 6 5 8...
Page 40: ...4 Measuring with the NBM 520 34 NBM 520 Narda...
Page 52: ...6 Instrument maintenance 46 NBM 520 Narda...
Page 56: ...7 Specifications 50 NBM 520 Narda 7 6 CE Declaration of Conformity...
Page 60: ...8 Ordering information 54 NBM 520 Narda...
Page 63: ......