13 Measurement Data Memory
226
IDA
-
3106
Narda
13.7
Saving measured values automatically
You can also save results automatically in the data memory instead of
pressing the
SAVE
key each time you want to save a result. There are two
memory functions for this purpose:
• Event triggered saving
Saving of results starts as soon as the trigger condition is met and a
measurement is triggered. Automatic storing controlled by trigger events
is only possible in
Time Domain (Scope)
and
I⁄Q Analyzer
modes.
• Timer controlled saving
Storing of data is controlled by a start time and a duration. Timer-
controlled storing is only possible in
Spectrum
,
Level Meter
and
Multi
-
Channel Power
modes.
NOTE:
The internal memory of the IDA
-
3106 has the capacity to store large
quantities of data. However, reading out a large number of data sets from
the instrument to a PC can take a long time. Try not to fill the data memory
too much, and carry out regular data maintenance.
13.7.1 Event triggered storing of measured values
(Scope)
NOTE:
Automatic storing of measured values controlled by trigger events is only
possible in
Time Domain (Scope)
and
I⁄Q Analyzer
mode. You can find
information about this under
Saving measured values automatically
13.7.2 Time controlled storing of measured values
(Time Controlled Storing)
Main Menu • Memory Menu • Time Controlled Storing
NOTE:
Timer-controlled storing is only possible in
Spectrum
,
Level Meter
and
Multi
-
Channel Power
modes.
NOTE:
Time controlled storing in Scope mode is started from the operating mode
itself and not from the Main menu. You can find information about this under
Summary of Contents for IDA
Page 1: ...IDA 3106 Interference and Direction Analyzer Operating Manual IDA IDA 2...
Page 8: ...Contents VI IDA 3106 Narda...
Page 28: ...2 Important Safety Instructions 20 IDA 3106 Narda...
Page 134: ...6 Common Functions 126 IDA 3106 Narda...
Page 166: ...7 Direction Finding Mode 158 IDA 3106 Narda...
Page 188: ...9 Level Meter Mode 180 IDA 3106 Narda...
Page 206: ...10 Time Domain Scope Mode 198 IDA 3106 Narda...
Page 218: ...11 I Q Analyzer Mode 210 IDA 3106 Narda...
Page 238: ...13 Measurement Data Memory 230 IDA 3106 Narda...
Page 242: ...14 Maintenance Repairs Updates 234 IDA 3106 Narda...
Page 246: ...15 PC Software 238 IDA 3106 Narda...
Page 254: ...16 Remote Control 246 IDA 3106 Narda...
Page 269: ...17 4 Antenna handle and antennas Narda IDA 3106 261 17 4 7 Antenna factors typical...
Page 272: ...17 Specifications 264 IDA 3106 Narda...
Page 284: ...19 Glossary 276 IDA 3106 Narda...
Page 290: ...Index 282 IDA 3106 Narda...
Page 291: ......