Description 4-7
4.5 Standard
Standard section allows the user to select standard ICNIRP, IEEE, BGV
and EC limits, already included by the software installation, as well as
user’s limits which can be created through the “Make Limit” function
provided by this section.
Safety Standard:
This section
can be used to load a default standard limit
saved into the memory at the factory. Default limit list shows limits that are
compatible to the actual setting only. If Electric field was selected in the
Mode section, electric field limits only are displayed. No default limit is
shown in case mG unit was selected.
User Defined:
This section allows to select and activate a specific limit
among limits created and saved by the user. They can be Linear or Non
Linear (governed by a mathematic formula).
The
Edit Limit
button opens a window to edit linear limits and create new
ones: