Procedure
Hold the ESM “RadMan” at arm’s length.
Move the ESM “RadMan” slowly along the
waveguide at a relatively large distance
from it.
If a leak is suspected, slowly move the
ESM “RadMan” closer to the suspected
location.
Once the 100% alarm stage has been
reached, do not move the instrument any
closer to the waveguide.
Tip:
The distance to the location of the leak
should be a minimum of 3 cm,
otherwise the effects of the ESM
“RadMan” field sensors on the field to
be measured can no longer be ignored.
Tip:
To determine cable damage reliably,
move ESM “RadMan” slowly over the
cable being checked.