background image

PRESTO

ϩ

Control Unit

Foot Pedal

Plumbing Connector

Handpiece Stand

Pressure Gauge

Plumbing Hose

Unit Mounting Screws

Air Supply Connector

Regulator Knob

PRESTO Handpiece

Handpiese Tubing Clamps

Cartridge

Filter Inspection Window

Handpiece Tubing

Bur Push-in Tool

Wrench Flats

Handpiece Connector

Foot Pedal Connector

Cartridge Wrench

Chuck Release Ring

Tube Stopper

Hose Stopper

Rotor Shaft Nut Wrench

2. COMPONENT NAMES

3. SPECIFICATIONS

Speed

320,000min

-1

Recommended Air Pressure

0.2 

.25MPa

Handpiece Dimensions

ȭ

16.6 (D) X130 (H) mm

Handpiece Weight

71g

Control Unit Dimensions

W120 X D102 X H165 (mm)

Control Unit Weight

720g

Proper Air Pressure

40N

 / min

Inside Diameter

ȭ

1.6mm Bur Only

4. SYSTEM ASSEMBLY PROCEDURES

4 - 1  Mounting of the Control Unit

The Control Unit can be secured on a wall, on a work-bench, or under a work-bench.

When securing on a wall surface, etc., fix the Control Unit with unit fixing screws in the two screw holes on its back.

4 - 2  Installation of Plumbing Hose

Push the plumbing hose into the plumbing connector located at the regulator on 

the left-hand side of the Control Unit until it is securely set to make connection 

as shown in Fig. 1. Check if the hose is securely installed by tugging it after 

connection.

Connect the other end of the plumbing hose to the air line. At this time, use the 

attached connector if necessary.

CAUTION

Push the plumbing hose into the plumbing connector until it is securely set. Otherwise, air may leak.

Pushing the white ring, on the plumbing connector, gently remove the tube.

Plumbing connector

Plumbing Hose

White Ring

Plumbing connector

Plumbing Hose

White Ring

Fig. 1

4 - 3  Installation of foot Pedal

Loosen and remove the hose stoppers (two) from the foot pedal connectors on the 

back of the Control Unit, and insert them into the foot pedal hoses. At this time, 

insert the hose stoppers so that their screws come outside (Control Unit side). (Fig. 2) 

For wall-hanging use, pass the hoses through the hole shown in Fig. 2.

Insert the "1" - marked end of each hose according to the instructions on the 

label. After insertion of both hoses, securely tighten the hose stoppers.

Foot Pedal Connector

Hose Stopper

When using by Wall-hanging
way, this hole can also be used.

Foot Pedal Connector

Hose Stopper

When using by Wall-hanging
way, this hole can also be used.

Fig. 2

4 - 4  Installation of Handpiece

Loosen and remove the tube stopper from the handpiece 

connector on the right-hand side of the Control Unit, and insert 

it into the handpiece tubing. At this time, insert the tube stopper 

so that its screw comes outside (Control Unit side). (Fig. 3)

Insert the handpiece tubing into the handpiece connector, and 

securely tighten the tube stopper.

Insert the tube clamp ring into the other end of the handpiece 

tubing as shown in Fig. 4, and insert it into the handpiece 

tubing socket as shown in Fig. 5.

Handpiece Tubing

Handpiece Connector

Tube Stopper

Handpiece Tubing

Handpiece Connector

Tube Stopper

Fig. 3

Fig. 5

Tube Clamp Ring

Handpiece Connector

Tube Clamp Ring

Handpiece Connector

Fig. 4

4 - 5  Installation of Handpiesce Tubing Clamps

Locate a Handpiece Tubing Clamp (two pieces supplied) at a proper location to conveniently position the handpiece tubing.

5. OPERATION OF CONTROL UNIT

5 - 1  Set the Drive Air Pressure

Supply air, and adjust to 0.25MPa by pulling the Regulator Knob upward and turning.

When adjustment is made, push the knob down to lock. (Fig.6)

5 - 2  Operation

Stepping on the Foot Pedal starts rotation.

5 - 3  For Finishing Work

Close the main air valve.

Regulator Knob

Fig. 6

6. OPERATION OF HANDPIECE

As the chuck of the PRESTO handpiece is a push-in type, set the 

ȭ

1.6mm burs and push in as shown in Fig.7.

If the end of the bur is tapered push in using the provided bur-

inserting wrench as shown in Fig.8.

When removing the bur, turn the chuck release ring in the direction 

of the arrow in Fig.9.

Fig. 7

Chuck Release Ring

Fig. 9

Fig. 8

7. REPLACEMENT OF CARTRIDGE

Fit the cartridge wrench supplied with the handpiece to the flats on the handpiece 

nose. Turn the cartridge wrench as shown in (Fig.10) after it is loosened, unscrew 

the nose with fingers and remove from the handpiece.

Before inserting a new cartridge, clean the head interior.

Finally, securely tighten the cartridge by turning it in the reverse direction to 

removal.

Fig. 10

① ② ③ ④ ⑤ ⑥

23

21

22

Reviews: