NAG/IM/ETC 600/01/22
14
Check that earth connection for the instrument is
available and then plug the Power cord
Always operate this instrument at room temperature
between 15˚C and 40˚C. Allow sufficient air
circulation for the instrument by leaving at least 20
cm of clearance around the instrument.
Do not turn off the instrument at temperature higher
than 50˚C. This would create a hazardous situation.
Select the set-point to room temperature and allow
the Instrument to cool before it turning OFF.
Do not
use the instrument if the cooling fan located
at the bottom of
the
Instrument is out of order.
Ensure
a free supply of air to the fan grill located at
the bottom of the Instrument.
STORING AND TRANSPORTING THE CALIBRATOR :
The following guidelines should always be observed
when storing and transporting the calibrator. This
will ensure that the calibrator remain in good
working order for longer period.