Naggura SWOP S301 Instruction Manual Download Page 5

Instruction 

Manual

5

Naggura treatment tables are only meant to be used by professional personnel in physiotherapy, rehabilitation, examination, mas-

sage and aesthetics fields.

It is the responsibility of the owner to ensure that training is given to all operators and third parties (like for example cleaning per-

sonnel) on safe use of the treatment table, and that they have read and understood all the operating and safety instructions. 

Only trained operators may use the treatment bed. Please keep the manual in a place where it is easily accessible to all users, and 

make it an important part of your employee training program.

6. Safety instructions

Responsibility of the owner

- Place the table in a space with free access from the 4 sides. Do not place the table against the wall or against any obstacles. 

- Do not use the table in a high-humidity environment or in so-called “wet rooms” (hydrotherapy spaces). Any prolonged exposure to 

high moisture conditions may affect the electrical components and may result in failure.

- Check the technical data and electrical parameters to ensure that the supply voltage meets the technical specifications of the 

table. If not, contact your supplier before connecting the device.

- Make sure that the unit is correctly plugged into an earth intake, following the local and national requirements and standards.

- Make sure that the power cable is correctly and safely placed  in such a way that it cannot get trapped  between any mechanism. 

- Pay special attention that the supply cable does not get pinched under the feet of the table, (e.g. after having moved the table on 

its wheels). 

- If the table accidentially gets wet, unplug it immediatelly.

- Always disconnect the bed from the power supply before inspection. Children are not allowed to use the bed or be near the equip-

ment while it is on use. Do not leave children unsupervised near the table.

Safety on installation and electricity

Naggura products are produced and marked in conformity with the directives hereby stated :

Low voltage : 2014/35/CE

Electromagnetic compatibility : 2014/30/CE

These products are manufactured according to the essential requirements established in the European Medical Devices Directive:

IEC 60601-1:2005 , IEC 60601-1:2005/AMD1:2012

EN 60601-1-2:2015 , IEC 60601-1-6:2010

IEC 60601-1-6:2010/AMD1:2013 , ISO 10993-10:2010

ISO 10993-5:2009

The Swop line has a guarantee

 of two (2) years in all parts.  The manufacturer will not be held responsible for the results of mainte-

nance or repairs performed by unauthorised persons. 

For further information on guarantee terms, please contact your distributor.

5. EC Declaration of conformity

Summary of Contents for SWOP S301

Page 1: ...Swop S301 SWOP S301 Treatment Table for Medical Examination NG 2021 S301 1 Instruction Manual EN VERSION 1 0 ...

Page 2: ...iability 5 EC Declaration of conformity 6 Safety instructions 7 Dimensions 8 Features Operation Wheels Levelling feet Gas piston lever Breather hole and plug Handlebar support Hand controller Foot controller Paper roll holder Interchangeable front panels LED lighting Cromo Mood Bamboo sound system 9 Maintenance and cleaning 10 Troubleshooting 11 Service ...

Page 3: ... will more than fulfil your most demanding work requirements NAGGURA s experts are in the lead of a project created with great devotion Every single detail has been cared to join technology performance and design altogether in a single table Our aim is designing and manufacturing INTUITIVE DYNAMIC and HIGH TECH tables of a DARING and UNIQUE design Thanks for your choice and start enjoying the NAGG...

Page 4: ...nts Naggura company is not liable for incidental or consequential damage to property or business 3 Warranty Under any circumstances the producer shall be liable for any indirect special incidental or consequential damage incurred by the user the patient or the table itself as a result of a A wrong diagnosis b Incorrect use of the table or any of its accessories c The wrong interpretation or non ob...

Page 5: ... Make sure that the unit is correctly plugged into an earth intake following the local and national requirements and standards Make sure that the power cable is correctly and safely placed in such a way that it cannot get trapped between any mechanism Pay special attention that the supply cable does not get pinched under the feet of the table e g after having moved the table on its wheels If the t...

Page 6: ...his purpose Do not sit on the head section or end of the leg section of the table The head section and armrests are not suitable to sit on Patients are not to be left unattended Never leave the table on its wheels during the treatment Do not exceed the maximum load capacity of the table Place the table on a stable and horizontal surface Level it if necessary with the levelling feet under the table...

Page 7: ...Instruction Manual 7 Swop S301 118 cm 73 cm 195 cm 51 92 cm 68 cm 40 cm 6 5 º Package dimensions cm 7 Product dimensions ...

Page 8: ...Instruction Manual 8 Handling and transport ...

Page 9: ... Manual 9 28 cm cm Voltage and frequency input shall be 220 240V 50 60Hz Plug the table cable into the socket as shown below 6 Operation Power input Lateral displacement cm Power cord length cm Product dimensions ...

Page 10: ...ever wheels 3 Gas piston lever 4 Breather hole and plug 5 Handlebar support 6 Hand controller 7 Paper roll holder 8 Interchangeable front panels 9 LED lighting Cromo Mood 10 Bamboo sound system 11 Foot controller Optional Operation of Swop S301 ...

Page 11: ...he table to have easy access when its height has to be adjusted Operation Hand controller Foot controller Down Up A foot controller is available as an option and it can be ordered separately To lift the table simply press the UP button on the foot switch Once the desired height is achieved realease the button To lower the table do the same operation with the DOWN button In order to use both the ha...

Page 12: ...mperature 0ºC 40ºC Relative humidity 30 90 Loading capacity 175 Kg Specifications Connecting the foot controller Locate the control box inside the table base frame Take the safety piece that locks the conectors into each plug and disconnect the cable HS Connect the branched cable in the plug named HS Place again the saftey piece to lock the connectors Now you can connect the handcontroller and the...

Page 13: ...our mood or however you feel today The exchangeable panels are 2 units one for one side of the table base and the other for its opposite The available panels are made of different materials metal wood and pmma and multiple colours You can visit our website to discover the full range of possibilities Changing the panels is so easy Simply unscrew the 2 screws that subject the panel with your fingers...

Page 14: ...Mood is installed an additional hand controller is provided to select your beloved colour to adjust the intensity of light or to change the lighting mode LED lighting Cromo Mood If you purchased the Bamboo sound system your swop table arrives with an awesome Bluetooth speaker we call it the Bam boo sound system The speaker is lodged inside a bamboo case with bevelled base You can place the speaker...

Page 15: ... cloth soaked in a mix of water with neutral diluted soap and rinse Dry with a clean cloth Maintenance Naggura tables have been designed for an intensive use during long time thanks to their stability reliability and durability However we advise to keep the table in perfect conditions and make a correct use of it Unplug the table if you will not use it for a long period of time Environment informa...

Page 16: ...nd gas lift Tha back rest is blocked it can t be lowerer nor raised up The gas lift activating pin does not activate the lever Remove the lower side of the gas lift and tighten it screw it 2 turns until you see that the pin can activate the lever Proceed the same operation with the second gas lift The up and down motor does not work Check if the power cable is disconnected either from the wall soc...

Page 17: ...Instruction Manual 17 Our products are subject to continuous technical improvements Naggura reserves the right to change or modify the product and contents without prior notice ...

Page 18: ...www naggura com Pol Ind La Valldan Camí de Garreta 92 08600 Berga Barcelona Spain T 34 93 824 90 20 E info naggura com ...

Reviews: