2. Safety Thermostat   

 

This is the safety device which prevents overheating caused by   

the product's malfunction.   

Please set the temperature at least 5 ~ 10

 

higher than set   

temperature. 

 

 

■ Using at a high temperature: Set SAFETY dial at higher than the set temperature. 

 

■  e.g.)  If  you  use  the  setting  temperature  37

,  set  SAFETY  S/W  more  than  about  42

.  You  can 

identify its activation point by lighting of pilot lamp. At set point, turn the safety dial and recognize 

its deviation. Considering the deviation, set the safety 5

 

higher than set point. 

 

■  It  is  set  at  42

 

when  the  product  is  released  from  the  factory,  but  it  may  be  changed  during 

transportation of the product, so please check it. 

 

■ SAFETY S/W is the safety device that prevents over-heating caused by malfunction of temperature 

control. 

 

When the SAFETY S/W is activated with visual alarm on pilot lamp(marked with safety), 

safe thermostat cut the heater automatically to prevent overheating.     

If temperature is not able to reach to set value, please check the safety dial is properly set.   

(at least 5

 

higher than set value). 

 

3. Service Contact   

 

     

If you need technical service support, please contact your local dealer or   

international sales team of N-BIOTEK.   

 

N-BIOTEK, Inc. 

402-803 Techno-Park, 655 Pyeongcheon-ro, WonMi-Gu, 

Bucheon-Si, GyeongGi-Do, KOREA (ZIP CODE: 14502)

 

Tel: (+82)32-321-2100 / Fax(+82)32-328-2372

 

E-mail

[email protected]

 

Summary of Contents for NB-201

Page 1: ...GENERAL INCUBATOR MODEL NB 201 C NB 201L OPERATING MANUAL...

Page 2: ...is proved to be defective by wrong workmanship after N BIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of installation or 1 and Half year from the date of shipm...

Page 3: ...pressed by back of the product A space between the product and the plug must be 20cm at least The power outlet must be only for this product Using various products simultaneously can cause a fire Clea...

Page 4: ...a hot place or a place that is near an electric heat The proper operation temperature is 17 30 Do not put inflammable substances near the product It can cause a fire CAUTION The product requires the...

Page 5: ...Do not disassemble the shaker during the product is operating CAUTION Do not wash the product with excessive quantity of water thinner benzene and Petroleum It can cause an electric leakage and malfun...

Page 6: ...electric shock an electric leakage and a fire can be occurred without an electric leakage breaker Do not ground to these places Gas Pipe water pipe pipe lighting rod telephone wire etc Wrong ground co...

Page 7: ...reproducibility are excellent by using Microprocessor Controller Safety Switch is equipped to prevent overshooting Inner glass door allows you not to open door for your observation inside Peltier NB2...

Page 8: ...han 17 Hence the minimum set up temperature is 17 when incubator is placed and used at 22 or lower Room Temperature d Press SET key again after putting the desired value SAVE is shown up on the LED sc...

Page 9: ...the factory but it may be changed during transportation of the product so please check it SAFETY S W is the safety device that prevents over heating caused by malfunction of temperature control When...

Reviews: