8. Humidity   

Mini CO

  incubator adopts natural humidity system by water container as below image. 

Please fill up distilled water up to the water line of water tray and keep lid closed for proper humidity level 

(at about 80%) during operation of incubator.   

Place the water container be close to back of chamber so the container is more exposed to heated air 

directly from water pass.   

No water is recommendable when carrying this incubator.   

In case that humidity is necessary during carriage, ensure the water container 

to be securely held at bottom of chamber(like using sticky mat or tape) and    wet 

sponge  or  sterilized  gauze  is  recommendable  rather  than  filling  water.  In  this 

situation,  humidity  may  not  be  reaching  to  maximum  or  normal  humidity  like 

normally max humidity up to 80%.       

 

9. Alarm 

 

Turn alarm system ON by pressing alarm button for about 5 seconds. When switching alarm on or off, 

a short alarm will come out. When alarm system is on, LED lamp is OFF. When alarm system is off, 

LED lamp of alarm is lighting to indicate that alarm system is disarmed.     

In order to activate alarm system, it should maintain 

±2

°

C,

 

± 

1%(CO2) from set point for 3 minutes. 

This course is recognised as stabilizing process. 

After stabilizing process, alarm system will be armed and alarm event is as below. 

 

Alarm Activation 

 

Temperature : (bird singing) 

- If it stays out of 

±1℃ 

from

 

set

 

point  for 8~9 minutes, it will give you an alarm. 

- Alarm will automatically stop once temperature is recovered into tolerance range. 

- Pressing mute button once will give 10 minutes delay.   

 

 

CO2 : (bird singing) 

- If it stays out of 

± 

1%

 

from

 

set

 

point  for 8~9 minutes, it will give you an alarm.   

- Alarm will automatically stop once CO2 is recovered into tolerance range. 

- Pressing mute button once will give 10 minutes delay.   

 

Door Open : 

  - Alarm comes out 30 seconds after door opening. 

  And no alarm come out if door is closed within 30 seconds. 

  - If door is closed during alarm, alarm will stop immediately.   

  - Pressing mute button in the situation of door open will give 3 minutes alarm delay. 

Summary of Contents for MINI-CELL NB-203M

Page 1: ...CO2 Incubator Operation Manual Model MINI CELL NB 203M...

Page 2: ...TENTS PAGE 1 WARRANTY 3 2 GENERAL INFORMATION ON PRECAUTION 4 3 TRANSPORATION STOARGE AND LOCATION OF INSTALLATION 6 4 PREREQUISITE AND CONFIGURATION 8 5 FEATURE AND SPECIFICATION 11 6 CONTROL PANEL 1...

Page 3: ...nty service will be provided once the unit is proved to be defective by wrong workmanship after N BIOTEK or reliable distributor s examination 2 The warranty period is 1 year from date of installation...

Page 4: ...rious products simultaneously can cause a fire Clean the power plug with a dry towel and connect it properly Foreign substances or unsafe connection can cause a fire Do not bend the power cable hardly...

Page 5: ...kage and a fire 3 Precaution for use You must not disassemble fix and remodel the product by yourself You can damage the product to a fire and malfunction or get a property loss as well as experimenta...

Page 6: ...nsportation DO NOT try to slide or tilt the unit Permissible ambient temperature range for transport 10 to 60 2 Storage Do not keep it at place in High Humidity Permissible ambient humidity max 70 sto...

Page 7: ...ell ventilated location If the ground is not flat it can cause an excessive vibration of the product When you move the product hold the door and other movable parts of the product with a tape When the...

Page 8: ...bator 3 Uneven ground or table 4 The place where is being vibrated Cleaning before use Before conducting cell culture it is recommended to clean up entire chamber and shelves and water tray by using d...

Page 9: ...1 Front Side Handle for carrying Control Panel Door Handle Figure 2 Inner Side and Rear Side IR CO Sensor Door Sealing Gasket Stainless Steel Panel Stainless Steel Shelf Water Container Circulation Fa...

Page 10: ...on slip rubber Note The tube diameter of CO2 inlet port is 4 x 2mm total diameter 4mm hole 2mm A connecting tube for CO2 inlet port is provided as well as normal size of tube 6 x 4mm which is for conn...

Page 11: ...tion Item Specification Chamber Volume 15 2 liter Temperature Range 15 C 45 C at Ambient 25 C Temperature Accuracy 0 25 C at 37 C Temperature Control Micom CO2 Range 0 20 CO2 Sensor Dual Beam IR senso...

Page 12: ...L LED Lamp RED Heating BLUE Peltier Cooling 3 ALARM ON OFF LED Lamp No LED Lamp ON RED OFF 4 Temperature LED Display 5 Temperature Setting Button 6 Adjustment Button Up Down 7 CO Setting Button 8 Cali...

Page 13: ...the level in all directions side by side front to back and ground 2 Connect the Power Prior to connect the power plug make sure that the POWER S W is OFF 3 Connect the CO2 Gas Check whether CO2 gas is...

Page 14: ...is too high it causes malfunction of the CO2 control Therefore it is highly recommendable to find a right level of pressure by user 4 Power Switch On Turn on the power switch in line of power cable T...

Page 15: ...in which takes about more than 2 hours You might want to perform this stabilization process at night before home Please note that low deviation range such as 0 1 0 3 may not be corrected precisely by...

Page 16: ...is 5 and Display shows 4 then press up 1 c Fifth Calibration Setting Heating Control NOTE Channel 5 is to set heating control point and is designed to prevent a significant temperature overshoot This...

Page 17: ...ll come out When alarm system is on LED lamp is OFF When alarm system is off LED lamp of alarm is lighting to indicate that alarm system is disarmed In order to activate alarm system it should maintai...

Reviews: