background image

 

7

 

 

 

6.

 

The Nitrate sensor needs to be conditioned after dry storage, see SENSOR CONDITIONING. 

II. STORAGE 

After  each  use 

ALWAYS

  rinse  the  Nitrate  sensor  with  clean  water  (preferably  DI,  RO,  or  distilled)  and  then  carefully  pat  the 

sensor with a clean soft lint-free cloth to remove any water drops.

 

1.

 

Short term storage (<=3 days): 
Keep the sensor in the protective cap filled half-full with prepared 100 ppm standard solution (DO NOT use standard solution 
prepared with NISB for storage). Store the PT6 in a cool place.  

2.

 

Long term storage (>3 days): 
a.

 

Rinse the Nitrate sensor with clean water (preferably DI, RO, or distilled) and then carefully pat the sensor with a clean 
soft lint-free cloth to remove any water drops. 

b.

 

Rinse the protective cap with clean water (preferably DI, RO, or distilled), pat with a clean soft lint-free cloth to remove 
any water drops. 

c.

 

Put the protective cap on the sensor and store the PT6 in a cool place.  

III. BATTERY REPLACEMENT  

The PT6 display has an indicator that depicts the PT6’s battery charge 
level.  When  the  charge  level  indicator  only  shows  one  bar, 
immediately replace the battery with a new N type battery.  
1.

 

In a CLEAN and DRY place unscrew the battery cap in a counter-
clockwise motion.  

2.

 

Slide the cap and battery housing out of the PT6.  

3.

 

Remove the depleted battery from its housing.  

4.

 

Insert  a  new  battery  into  the  battery  housing  oriented  with  the 
negative end touching the spring.  

5.

 

Align  the  groove  along  the  battery  housing  with  the  guide  bump 
inside the PT6 case and slide the battery housing back in.  

6.

 

Screw the battery cap back on in a clockwise direction. Do not over tighten. 

IV. REPLACING OR REINSTALLING THE NITRATE SENSOR 

 

Follow the instructions that come with your replacement sensor.   

 

Only remove/replace the Nitrate sensor in a CLEAN and DRY environment. 

To remove the Nitrate sensor:  
1.

 

Remove the protective cap by wiggling it side to side while you pull it off the PT6. 

2.

 

Make sure the PT6 (including the sensor) is clean and dry. 

3.

 

Loosen the battery tray (to allow pressure equalization). 

4.

 

Firmly grasp the sensor body and slowly pull the sensor out. DO NOT allow any dust or moisture into the body of the PT6. 

To install a new Nitrate sensor:  
1.

 

Line up the alignment tabs on the sensor with the alignment slots on the PT6 unit.  

2.

 

Gently push the sensor into position. 

3.

 

Tighten the battery cap. 

NOTE: 

The PT6 

MUST

 be recalibrated any time the Nitrate sensor is replaced. 

V.  TROUBLESHOOTING 
1. SENSOR CHECK 

Sensor Check verifies the slope of the recent calibration. If readings are not as expected, or measurements of calibration solutions 
after  calibration  are  not  within  ±10%  of  the  printed  value,  checking  the  slope  of  current  calibration  can  be  a  starting  point  for 
troubleshooting. To do a Sensor Check: 
1.

 

Press and release the push button to turn the PT6 ON. 

2.

 

Push and hold the push button to enter menu mode. 

3.

 

Release the button while “SEnSor 

” is displayed. The display will flash “SEnSor 

” and then indicate the mV value of the 

slope for current calibration.    

4.

 

A gain ranging between -50 to -62 mV indicates a normal calibration in the linear range. Values beyond that may indicate a 
calibration out of the linear range, an insufficiently conditioned sensor, or a worn membrane.  Most of the time the issue can 
be  resolved  by  conditioning  the  sensor  for  at  least  one  hour  and  re-doing  a  2-point  calibration.  If  the  problem  persists,  the 
sensor should be replaced.  

NOTE

: The gain range of -50 to -62 mV is based on a calibration temperature of 25˚C.  

Summary of Contents for Ultrapen PT6

Page 1: ...he sensor submerged 5 In HOLD mode Real time readings are displayed until the LED turns solid ON The display holds the final reading briefly then the PT6 turns OFF 6 In LIVE Mode The PT6 will display...

Page 2: ...mV displays millivolts and Temperature ESC 6 Release the push button when desired parameter type is displayed SAVEd will display then the PT6 will power OFF II TEMPERATURE UNIT SELECTION The PT6 allow...

Page 3: ...re in Nitrate NO3 To select the optimal Standard Solutions and to verify calibration see the CONVERSION TABLE below CONVERSION TABLE Standard Solution Value in NO3 N Value in NO3 N1SOL 1 ppm 4 4 ppm N...

Page 4: ...cts an error in the gain slope it will display Error If this happens clean and condition the sensor then verify you are using approved and properly prepared calibration solutions and recalibrate NOTE...

Page 5: ...s hold the sensor steady during measurement In HOLD mode real time readings are displayed until the LED is ON solid In LIVE mode real time readings are displayed until the PT6 turns OFF 90 sec in HOLD...

Page 6: ...y using concentrated standard solutions to increase the nitrate concentration in the sample by a known value optimally double the measured value and then compare the value with the increase in Pen rea...

Page 7: ...LLING THE NITRATE SENSOR Follow the instructions that come with your replacement sensor Only remove replace the Nitrate sensor in a CLEAN and DRY environment To remove the Nitrate sensor 1 Remove the...

Page 8: ...ngs ca gov 1 Accuracy of measuring calibration solutions after Temperature Controlled 2 Point Calibration with one order of concentration difference 2 For extended life of the sensor store the PT6 in...

Reviews: