Myron L ULTRAPEN PT5 Quick Start Manual Download Page 2

Myron L

®

 Company – 

ULTRAPEN™ PT5

 – Quick Start Guide

 

Dissolved Oxygen (DO) and Temperature Pen 

PT5QSI  08-17 

 

2   

V. AIR CALIBRATION (continued)

 

6. 

Push and hold the push button to enter menu mode. 

7. 

Release the push button while “CAL” is displayed.  The display will 
alternate between “CAL” and “PUSHnHLD”.  

8. 

Push and hold the button then release while “Air CAL” is displayed.  

9. 

The PT5 will perform the calibration.  When calibration is complete, the 
display will indicate “SAVEd” then turn OFF. 

10. 

To verify calibration follow steps 1 thru 5 above (taking a 

measurement) at 25 ºC, the readings should be 104% ±2%. 

NOTE:

 The PT5 

MUST

 be recalibrated any time: the DO sensor is 

replaced; the DO membrane or the electrolyte is replaced, or; the Zinc 
electrode is cleaned 

VI. DO Measurement 

 

CAUTION

:  For  best  results  when  measuring  solution  at  the  extremes  of 

the  specified  temperature  or  DO  range 

ALWAYS

  perform  a 

TEMPERATURE  CONTROLLED,  FULL  CALIBRATION  as  described  in 
the  full  PT5  Operation  Manual.    Allow  the  PT5  to  equilibrate  by 
submerging the sensor in the sample solution for 1 minute prior to taking 
a measurement.  

 

ALWAYS make sure that the DO Sensor is completely submerged in 
the solution being tested. 

 

If the expected measurement is < 3 ppm it is particularly important to 
perform the ZERO calibration described in the full PT5 Operation 
Manual. 

 

1. 

Push and release the push button to turn the PT5 ON. 

2. 

Dip PT5 sensor in solution to be tested and swirl. 

3. 

Continue to swirl the PT5 in solution while measurement is taking 
place.

 

The following table explains what the LED Indicator light signals indicate 
and gives the duration of each signal. 

NOTE

:  You  can  turn  OFF  the  unit  at  any  time  during  live  measurement 

with a simple press and release of the push button. 
 

LED Signal

 

Action

 

Duration

 

Rapid Flashing

 

Dip pen in solution and swirl. 

6 sec 

Slow Flashing

 

Measurement in process; continue 
to swirl. 
In 

HOLD

 mode real-time readings 

are displayed until the LED is ON 
solid. 
In 

LIVE

 mode real-time readings 

are displayed until the PT5 turns 
OFF. 

 
20 -120 sec in 
Hold mode 
 
Up to 5 min in 
LIVE mode 

Solid Light  

(Hold mode only

)

 

Measurement is complete. Values 
are displayed until the PT5 turns 
OFF. 

6 sec 

 

MAINTENANCE 

I. Routine Maintenance  

1. 

After each use 

ALWAYS

 rinse the DO sensor with clean water (preferably DI, RO, or 

distilled water), then carefully blot the sensor membrane with a soft, clean, lint-free 
cloth to remove any water drops.  

2. ALWAYS

 replace the hydration cap on the DO sensor after each use.  

3. 

To replace the membrane cap and/or refill electrolyte solution follow the SENSOR 
SETUP instructions. 

4. 

Do not drop, throw or otherwise strike the PT5. This voids the warranty.  

5. 

Do not store the PT5 in a location where the ambient temperatures exceed its 
specified Operating/Storage Temperature limits.

 

NOTE: 

For instructions on cleaning the DO sensor, performing a sensor check, 

troubleshooting, or storing the PT5, download the full PT5 Operation Manual from 
www.myronl.com.

 

II. Battery Replacement

 

1. 

In a 

CLEAN DRY

 place unscrew the battery cap in a counter-clockwise direction.  

2. 

Slide the cap and battery housing out of the PT5.  

3. 

Remove the depleted battery from its housing.  

4. 

Insert a new battery into the battery housing oriented with the negative end touching the spring.  

5. 

Align the groove along the battery housing with the guide bump inside the PT5 case and slide the battery housing back in.  

6. 

Screw the battery cap back on in a clockwise direction. Do not over tighten. 

SPECIFICATIONS 

Measurement Ranges 

DO Concentration: 0-20 ppm, DO Saturation 0-200%   
Temperature: 0-71°C / 32-160°F 

Accuracy

1

 

Concentration: 

±0 .2ppm or ± 2% of the reading, whichever is greater.  

Saturation:

 0.0% to 

 100.0%: ± 20 counts;  

>100%:  ± 2% of the reading.   

Temperature: 

± 0.1°C

 

Time to Reading 
Stabilization 

10 - 180 seconds 

Operating / Storage 
Temp 

0-55ºC / 32-131ºF 

Resolution 

DO Concentration: 0.01 ppm, DO Saturation 0.1% 
Temperature 0.1ºC/F 

Altitude Adjustment 

-2,000 ft. to 15,000 ft. in increments of 100 ft. 

Salinity Adjustment 

0 - 50 ppt in increments of 1 ppt 

Power Consumption 

Active Mode: 33 - 140 mA, Sleep Mode: 2 µA 

Battery Type 

N type, Alkaline, 1.5 V 

Temp Compensation 

Automatic Temperature Compensation for Membrane 
permeability 

Calibration Methods 

1 or 2 Point; 0 ppm DO Calibration Solution; Air or Water 

Physical Dimensions / 
Weight 

17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D;   
Weight: 54g or 1.94oz 

Case Material:  

Anodized Aircraft Aluminum with Protective Coating 

Water Resistance 

IP67 and NEMA 6 

EN61236-1: 2006 -Annex A: 2008; 1999/5/EC Annex III: Electrostatic discharge to  
case of instrument may cause PT5 to spontaneously power ON. In this case, the  
PT5 will power OFF after several seconds.

 

Accuracy after Temperature Controlled, Full Calibration. For this procedure and for more detailed accuracy specs download the full PT5 Operation Manual from  www.myronl.com. 

 

MYRON L

®

 COMPANY

 

2450 Impala Drive   

!

   Carlsbad, CA 92010-7226   

!

   Phone: +1-760-438-2021 

E-Mail: Customer Service – [email protected]   

!

   Technical Support – [email protected] 

Website: www.myronl.com 

Reviews: