background image

Myron L

®

 Company – Ultrameter III

 9P Quick Start Guide 

 

Handheld Conductivity, Resistivity, TDS, pH, ORP, Hardness, Alkalinity, LSI, Free Chlorine Equivalent (FC

E™

), and Temperature Meter

 

 

 

 

9PQSI 06-19

 

This Quick Start Guide is for the Myron L

®

 Company Ultrameter lll Model 9P.  Download the full 9P Operation Manual (P/N 9POM) from www.myronl.com to get more 

detailed instructions for taking measurements (including Alkalinity, Hardness, LSI calculations, and FC

E

), storing or recalling measurements, changing instrument 

settings, measurement techniques, maintenance procedures, calibration procedures, Troubleshooting Guide and FACTORY CAL reset. 

 

 

     

QUICK REFERENCE INSTRUCTIONS

 

 

Press and release any one of these buttons to turn on the 9P and begin taking measurements.   

 

Press         to calculate LSI.  

 

Press and release the CAL/MCLR button to calibrate (CAL) the active measurement parameter or press and hold to clear displayed memory location (MCLR). 

 

Press and release to store (MS) a measurement or move upward on the list of stored measurements (Location 1 to 100) while in memory recall mode. 

 

Press and release to view the most recent stored measurements or move downward on the list of stored measurements (Location 100 to 1).  Also used to enter 
Settings Mode. 

 

 

9P

 

 FEATURES 

1.

 

TEMPERATURE SENSOR – 

Houses thermistor for making temperature measurements. 

 

2.  CONDUCTIVITY CELL

 – 

Contains electrodes that generate a flux field in defined area for all measurements except  

pH, ORP, and FC

E

.  

  

3.  pH/ORP SENSOR 

(user replaceable) – 

Measures test sample’s pH, ORP, and Free Chlorine Equivalent (FC

E

)

4.

 

pH/ORP SENSOR PROTECTIVE CAP

 – Snaps in place to protect the pH/ORP sensor when not in use and seals in 

the sensor storage solution. 

5.

 

LCD DISPLAY – 

Displays measurements, units of measure, active measurement mode, current solution mode  

 

setting, low battery warning (LOBATT), titration prompts, and other information (see below).

 

6.   MEASUREMENT KEYS

 – Press and release any one of these to turn on the 9P

 

and begin taking measurements. 

7.  FUNCTION KEYS

 – Press and release to store the current measurement (MS) or recall a stored measurement (MR).  

Also used to move between memory locations and to select various mode settings

8.  CAL/MEMORY CLEAR KEY

 – Use to calibrate (CAL) the active measurement parameter, enter or clear the contents  

of the currently displayed memory location (MCLR). 

9

BLUDOCK SYMBOL

 – Indicates that this Ultrameter III 9P is equipped with Myron L

®

 Company’s optional wireless  

transceiver for wireless data download. 

10

CELL EXTENDER – 

Extends cell cup and holds sample and reagents during Alkalinity and Hardness measurements. 

11.  PLUNGER

 – Used to create proper cell cup volume of liquid during Alkalinity and Hardness measurements. 

12

CELL EXTENDER CAP

 – Seals cell extender while mixing sample and reagents during Alkalinity and Hardness  

 

measurements. 

 

LCD DISPLAY 

1.  MAIN VALUE DISPLAY 

– Displays the current measurement value and settings menu information. During Alkalinity & Hardness titrations, prompts appear here at 

various stages of the process. 

2.  UNITS OF MEASURE ICONS

 – Appropriate icon is displayed to indicate the unit of measure for the current measurement. 

3.

 

TEMPERATURE UNITS OF MEASURE

 – Displays °C (Celsius) or °F (Fahrenheit) based on setting selected for 

Temperature readings. 

4.

 

MEASUREMENT ICONS

 – Appropriate icon is displayed to indicate what type of measurement is being made. 

5.

 

BUFFER ICON 

– Appears when the 9P is in pH Calibration mode to indicate that the instrument is expecting a pH 

buffer solution. 

6.

 

TEMPERATURE DISPLAY

– Displays the temperature of the sample solution, memory location number  

(1 to 100) or buffer during pH calibration. 

7.

 

SOLUTION MODE ICONS

 – Appropriate icon is displayed to indicate the current solution temperature compensation 

mode setting for Conductivity, Resistivity and TDS measurements. 

8.

 

LOBATT ICON

 – Appears when the 9P ’s battery requires replacement.    

9.  CAL ICON – 

Is displayed

 

when the 9P is in Calibration mode. 

10.

 

MEMORY ICON

 – Is displayed to indicate that the values and icons being displayed are for a measurement stored in the 9P

 

’s memory (NOT a live measurement). 

 

OPERATING INSTRUCTIONS 

DEFAULT SETTINGS:  

 

Temperature Units:

 °C  (Temperature is displayed in Degrees Celsius) 

 

Solution Modes: 

Conductivity – KCl, Resistivity – NaCl, TDS – 442™ 

MEASUREMENT SETUP: 

Before you take a reading, make sure the 9P is clean, 

calibrated, and if measuring COND or TDS, that the desired solution mode has 
been selected (See Section II, Solution Mode Selection, below). The sample 
solution must also be within the specified range. 
For pH, ORP and FC

E

 measurements both the pH/ORP sensor well and 

conductivity cell cup must be rinsed and filled with sample solution. 
For greatest accuracy, resistivity, ORP, and FC

E

 measurements should be taken 

from a continuous flow of sample solution. 

I. MEASUREMENT 

1.

  Rinse the sensor well and/or cell cup 3 times with sample to be measured. 

2.

  Fill the sensor well and/or cell cup with fresh test sample. 

3.

  Press the appropriate Measurement Key: 

 

 

COND: Measures Conductivity 

 

TDS: Measures Total Dissolved Solids 

 

RES: Measures Resistivity  

 

displays as PPT of NaCl 

ORP/Fr Chl: Measures ORP or Free Chlorine Equivalent (FC

E

)* 

 

pH: Measures pH 

Press one of these to begin an Alkalinity* or Hardness* titration. 

Press to calculate LSI*. Uses stored or hypothetical Alkalinity and Hardness 
data.  

* For complete instructions on performing these measurements, download the full 

Ultrameter III 9P Operation Manual.  

 

ORP 

Fr Chl 

 

Reviews: