![Myron L POOLPRO PS9TK Operation Manual Download Page 54](http://html.mh-extra.com/html/myron-l/poolpro-ps9tk/poolpro-ps9tk_operation-manual_1863025054.webp)
50
Xvii. TROubLEShOOTing chaRT
Symptom
Possible cause
corrective action
No
display
, even though
measurement key pressed
Battery weak or not connected.
Check connections or replace battery. Ref. Battery Replacement, pg. 47.
Inaccurate
ph
readings
1. pH calibration needed. Ref. pH Cal.,
pg. 33.
2. Cross-contamination from residual pH
buffers or samples in sensor well.
3. Calibration with expired pH buffers.
1. Recalibrate instrument.
2. Thoroughly rinse sensor well.
3. Recalibrate using fresh buffers. Ref. pH Buffer Solutions, pg. 53.
No response to
ph
changes
Sensor bulb is cracked or an
electromechanical short caused by an
internal crack.
Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 54.
Will not adjust down to
ph
7
pH/ORP sensor has lost KCl.
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 48) and recalibrate. If no
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 58).
ph
readings drift or respond slowly to
changes in buffers/samples or “
Fac
” is
displayed repeatedly
1. Temporary condition due to memory” of
solution in pH sensor well for long periods.
2. Bulb dirty or dried out.
3. Reference junction clogged or coated.
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 48) and recalibrate. If no
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 58).
Unstable
conductivity/TDS/
Min/SaLT
readings
Dirty electrodes.
Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 48.
Unable to calibrate
conductivity/TDS/
Min/SaLT
Film or deposits on electrodes
.
Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 48.
alkalinity
or
hardness
titration with
calibration
solution value not as
stated on bottle
1. Method error.
2. Dirty Cell Cup.
3. Dirty pipette.
4. Faulty pipette.
5. Contaminated calibration solution.
1. Repeat titration.
2. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 48.
3. Clean pipette per manufacturer’s instructions.
4. Use a new pipette.
5. Use fresh calibration solution. Be sure to use a new tip for each type of solution
measured.
Low
ORP
Reading
Slow or no response to
ORP
changes
ORP platinum electrode is dirty.
Check the ORP sensor functioning. Take an ORP reading of Myron L pH/ORP
Sensor Storage Solution (ref. pH Sensor Storage Solution, pg. 53). If the reading is
outside the range of 350-400 mV, clean ONLY the platinum ORP
electrode with Myron L ORP Conditioner solution-soaked cotton swab (ref. ORP Sen
-
sor Conditioner Solution, pg. 53), being careful not to touch the swab to the glass bulb
of the pH sensor.
FC
E
responds very slowly or returns an
atypically high Predictive
ORP
value
1. Dirty platinum electrode (see above).
2. ORP sensor memory/battery effect.
Some ORP sensors exhibit a residual
charge when measuring LOW Free
Chlorine concentrations soon after
measuring a HIgH Free Chlorine
concentration.
1. Rinse the pH/ORP sensor well briefly with a small amount of ORP Sensor
Conditioner Solution.
DO nOT
leave the conditioning solution in the sensor well
for more than 10 seconds.
2. Rinse the pH/ORP sensor 3 times with Sensor Storage Solution.
3. Fill the sensor well with Sensor Storage Solution and let rest for 5 minutes.