Whisper
JECT
™
IMPORTANT:
Please read these instructions
prior to administering your first injection, even if you have used similar autoinjector devices.
INSTRUCTIONS FOR USE
Rx only
Page 1: ...WhisperJECT IMPORTANT Please read these instructions prior to administering your first injection even if you have used similar autoinjector devices INSTRUCTIONS FOR USE Rx only ...
Page 2: ...xed needle of 27G to 29G The WhisperJECTTM Autoinjector is a reusable injection device for the subcutaneous injection of FDA approved drugs For instructions regarding the use of the drug refer to the drug patient information booklet provided with the drug product If the device does not function do not use For more information call Mylan Pharmaceuticals Inc at 1 877 446 3679 1 877 4 INFO RX Made in...
Page 3: ...e the WhisperJECT Autoinjector without a 1 0 mL BD pre filled glass syringe or with an empty or used 1 0 mL glass syringe This could damage the WhisperJECT Autoinjector After inserting the 1 0 mL BD pre filled glass syringe DO NOT attempt to remove the Orange Needle Cap Remover unless the WhisperJECT Autoinjector is fully assembled DO NOT insert anything other than the Orange Needle Cap Remover an...
Page 4: ...hite Alignment Arrows Needle Depth Indicator Needle Depth Adjuster Handle 1 0 mL BD pre filled glass syringe in Protective Blister Packaging If you need help or have any questions please call 1 877 446 3679 1 877 4 INFO RX Syringe Housing Needle Depth Setting Value Orange Needle Cap Remover Plunger Rod Glass Rim Needle Cap Plunger 10 6 7 ...
Page 5: ...to handling supplies Gather the following supplies WhisperJECT Autoinjector in Protective Case 1 0 mL BD pre filled Glass Syringe not included with device Sharps Container not included with device Clean damp cloth for device cleaning post injection not included with device Alcohol swab for skin not included with device 10 1 0 mL BD pre filled Glass Syringe 8 9 ...
Page 6: ...e Cap Remover into the Syringe Housing Note The Orange Needle Cap Remover does not click into place DO NOT continue to Step 2 until the Orange Needle Cap Remover is inserted 1A Remove the Handle Orange Needle Cap Remover and Syringe Housing from the Protective Case 10 11 ...
Page 7: ...using for stability push the Handle down onto the Syringe Housing until you feel and hear a CLICK This signals that the device has locked into position The Viewing Window should be blank no green hash mark or check mark visible 2B Place the Syringe Housing on a clean flat surface with the Orange Needle Cap Remover end up 2A Step 2 Reset the Autoinjector 12 13 ...
Page 8: ...esult in a deeper injection Check with your healthcare provider to determine your correct needle depth setting DO NOT unscrew the white Needle Depth Adjuster from device Step 3 Set the Needle Depth Confirm the Autoinjector is reset When the Autoinjector is reset The white check mark inside a green box will no longer be visible and the white Plunger will occupy about half of the Viewing Window The ...
Page 9: ...emove the Orange Needle Cap Remover until syringe is inserted and device reassembled 4C WARNING 4A 4B Step 4 Insert 1 0 mL BD pre filled Glass Syringe into Syringe Housing Remove the 1 0 mL BD pre filled Glass Syringe from the protective blister packaging 1 0 mL BD pre filled Glass Syringe is not included with the WhisperJECT Kit DO NOT remove the 1 0 mL BD pre filled Glass Syringe Needle Cap Plac...
Page 10: ...rochure for the allowable injection sites Clean injection site with alcohol swab Step 6 Select and Clean Injection Site Reassemble the device by aligning the white arrows and turning the Syringe Housing until it locks into the Handle You will hear a CLICK as the two parts lock together Once the Autoinjector is fully assembled you are ready to start the injection process 18 19 ...
Page 11: ...uture use DO NOT dispose of the Orange Needle Cap Remover in the trash DO NOT recap the syringe 7A Hold the Handle with one hand and the Orange Needle Cap Remover in your other hand and then pull the Orange Needle Cap Remover out of the Syringe Housing The grey pre filled Syringe Needle Cap will be removed DO NOT replace the Orange Needle Cap Remover onto the Syringe Housing This may damage the ne...
Page 12: ...e inside the window If you cannot see the Viewing Window slowly count to 10 seconds after pressing the Injection Button and hearing the first CLICK DO NOT lift up early 8A Gently place the Autoinjector straight on the injection site not at an angle Make sure you can comfortably reach the orange oval Injection Button If possible position so you can see the Viewing Window Gently press the Autoinject...
Page 13: ...p from the injection site Always confirm that the full dose was delivered by checking the Viewing Window If the full dose was delivered you will see a white check mark in a green box inside the window If you believe you did not receive a full dose immediately contact your healthcare provider WARNING DO NOT put your fingers near the needle end of the Autoinjector The Needle Depth Adjuster which cov...
Page 14: ...frigerator DO NOT store 1 0 mL BD pre filled glass syringe inside the device Step 12 Place the Autoinjector in Protective Case for Storing Wipe the external parts and the accessible areas of the Syringe Holder and Handle with a clean damp cloth after each use IMPORTANT DO NOT clean the Autoinjector using any other methods e g DO NOT immerse in water as this can damage the Autoinjector DO NOT clean...
Page 15: ...icals Inc WhisperJECT is available by prescription only WhisperJECT is a trademark of Mylan Pharmaceuticals Inc BD is a registered trademark of Becton Dickinson and Company For more information call1 877 446 3679 1 877 4 INFO RX or visit Mylan com Revised 4 2017 IFU WJECT R1 ...