MYERS MA-26X Manual Download Page 2

21 Washington Avenue

Scarborough, ME

21 Washington Avenue

Scarborough, ME

888.293.2121

[email protected]

www.myerstest.com

888.293.2121

[email protected]

www.myerstest.com

Safety Instructions

Review and understand all of the operating and safety instructions before using this unit. 

WARNING!! 

The calcium carbide reagent used with this test must be handled with great care. When exposed to 

water, calcium carbide produces highly flammable or explosive acetylene gas. 

• 

Users must understand and follow local regulations and calcium carbide Material Safety Data Sheet guidelines and 

procedures regarding transportation, storage, handling and disposal. 

• 

Testing must not be performed in confined spaces or in the vicinity of an open flame, lighted cigarettes, embers or 

any potential source of ignition. 

• 

When releasing the gas from the apparatus, make sure to direct it away from your face and body. 

• 

DO NOT store, dispose of, or place the calcium carbide reagent where it may come into contact with water, because 

it can produce explosive acetylene gas. 

• 

Any application not specifically covered in this manual will be considered improper use and will void the warranty. 

• 

Testing excessively wet material, adding water to the testing chamber, or improper use of the equipment could 

cause pressures to exceed the safe level for the apparatus. 

• 

Long sleeves, gloves and eye protections should be worn at all times when performing the test. 

Unpacking & Set-Up

NOTE: Each Aqua-Check is carefully inspected prior to packaging and shipment. Immediately unpack and 

inspect the apparatus for signs of damage once received. If any damage has occurred, report it immediately 

to the shipping company. 

The MA-26 Aqua-Check consists of: 

• 

Cast aluminum Sample Chamber Vessel with integral

• 

0-21% Pressure Gauge

• 

200 x 0.1g Electronic Balance for weighing of test specimens 

• 

(2) 1.25in (32mm) Steel Pulverizing Balls 

• 

Long-handled Reagent Scoop

• 

Large, coarse clean-out Brush 

• 

Small, fine-bristle brush 

• 

Sample cup 

• 

Heavy-duty plastic case 

Introduction

The Aqua-Check Moisture Tester is designed to measure the water content of soils, aggregates, and other 

non-reactive materials. The Aqua-Check consists of an aluminum sample chamber vessel with integral pressure 

gauge, electronic balance for sample weighing, two steel pulverizing balls, and assorted brushes and containers 

as required for proper completion of the test. 

Pulverized calcium carbide is used as a reagent, and reacts when combined with water in the sample material to 

form acetylene gas. The most accurate results are obtained when pressure gauge readings are correlated with 

data from over-dry methods. 

1.25in Steel Balls

Long-handled Reagent Scoop

Aqua-Check Pressure Gauge

Reviews: