Operation/Consistency Determination 

After molding the sample as directed by the application specification, center the sample under the 

plunger and bring the two in contact. Loosen the knurled by adjusted the screws on the plunger 

and indicator if adjustment is needed. Release the plunger within the allotted time after mixing and 

take a reading at the specified time. 

Use different percentages of water or fluids to make trial cement pastes or other specimens to 

obtain the target consistency. Use fresh material when making each specimen. Calculate and report 

the required mix values. Refer to the applicable agency test standard for complete requirements. 

Maintenance 

Frequent cleaning of the plunger to prevent sluggish or sticky movement is recommended. Dry 

wiping of the plunger with a link free cloth will usually suffice, isopropyl alcohol can be used if 

needed to remove gummy deposits. DO NOT apply any type of lubricant to the spindle. Clean 

the scale by using a dampened soft cloth with mild detergent. 

Note: Use of aromatic solvents could cause damage to the device. 