equipment's safety.
(26) When operating the equipment, start from the lowest energy level and gradually
increase it after the client gets used to it.
(27) You can use the handpieces separately, but you can use them together to achieve
a better effect.
(28) To avoid discomfort caused by overheating, please shun the air outlet when
operating the hot & cold hammer.
(29) You must keep the treated parts moist and avoid operating the instrument in dry
skin conditions.
(30) Use the equipment and train the operators in strict accordance with the
instructions specified in the manual.
(31) After the operation, clean the equipment with normal saline to ensure its
cleanliness and hygiene to extend its service life.
4. Troubleshooting
(1) I can't start the equipment, and the indicator lamp doesn't light up.
A. Make sure you plug the power cord into a good power socket.
B. Check whether the fuse tube in the back is loose or burnt out.
(2) The machine doesn’t have energy output.
A. Check whether the handpiece and the plug connecting to the machine body are
closely linked.
B. Please check whether the treated parts are cleaned. Grease or essential oil may
cause poor contact between the handpiece and the body, resulting in no output.
(3) The energy output become weak.
A. Check whether the handpiece and the plug connecting to the machine body are
closely linked.
B. Please check whether there is non-conductive grease on the handpiece, which
may cause bad contact, resulting in a weakened output.
C. Check whether the products in use are the adaptive ones specified by the
equipment.
(4) I can start the equipment, but an error message shows on the monitor.
A. Take out the power plug from the back, wait about one minute, reinsert the plug,
and restart the equipment.
B. If the solution mentioned above doesn’t work, please contact the distributor for
assistance.
Summary of Contents for MS-78S1
Page 1: ...User Manual MS 78S1 ...