7. Patients who are suffering from illness should consult with a doctor and gain
permission from the doctor before using.
8. Please turn off the power switch of the device, and ensure that the total power
supply of the device is turned off before the staff leaves after daily use, so as to
ensure the safety of the electrical products.
9. Use device or train device operators in strict accordance with instructions in the
user manual.
10. Start from the lowest energy level and slowly add up.
11. When using this device, the operating parts must be kept moist and dry skin
should be avoided.
12. Please refrain from operating around eyes, ears, skin lesion or parts that had
undergone plastic surgery.
13. Adjust suction subjecting to the customer’s comfort level when using diamond
dermabrasion handle.
14. Do not stay at one place for over 5 seconds to avoid scald or frostbite.
15. Remind customer to adjust breathing when operating on philtrum with hydro
spray jet.
16. Diamond dermabrasion, ultrasound, and RF handles can be used alone or
combined together for better beauty results.
17. Clean the instrument with normal saline after operation to ensure its cleanliness
and hygiene to prolong its service life.
18. Make sure to use rubber ring during diamond dermabrasion to avoid low or no
suction.
19. Make sure to user cotton filter to avoid damage caused by dirt in the tube.
20. Make sure skin is moist during diamond dermabrasion treatment for better
comfort level.
21. Distinguish inhale hole from exhale hole. Do not insert diamond dermabrasion
probe into the exhale hole, which will result in improper machine functioning.
22. Inform the customer of possible BIO experience before treatment to avoid
unexpected results.
5. Troubleshooting & Solutions
1. The instrument can't start and the key lamp on the back of the instrument doesn't
work?
A. Ensure that the power cord is connected to a working power outlet.
B. Ensure the fuse tube on the back of the instrument is not loose or being burnt.
2. No RF output of the instrument?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether the treatment area has been cleaned. Grease or oil essential
oil products may cause poor contact between the treatment handle of the instrument