membrane on the treated area and dry skin should be avoided.
17. Clean the instrument with normal saline after operation to ensure its cleanliness
and hygiene and prolong its service life.
18. The twisting knob is for controlling vacuum suction levels. By twisting, it adjusts
suction level of (vacuum cooling applicator).
19. Fixate the laser pads on the customer first, then turn on the machine and adjust
energy level; It is advised to turn the energy level from low to high slowly subjecting to
the customer’s comfort level.
20. Make sure the whole surface of the laser pad come in full contact with the skin,
not just a small patch of skin.
21. Lipo laser paddles can be used at separate occasions; Fixate it directly on fat part
and adjust energy; Each session should last around 20 minutes or so. Combine
massage, RF and 40K with laser pads for a better beauty result.
22. It is advised not to use vacuum fat-freezing and 40K, RF at the same time. When
conducting 40K and RF treatment, you need to wait for 2 weeks or so to conduct
vacuum fat-freezing treatment, or when conducting vacuum fat-freezing treatment,
you need to wait for around 28 days before you can conduct 40K and RF treatment.
4. Troubleshooting & Solutions
1. The instrument cannot be started, but the key light on the back of the instrument
does not work?
A. Make sure the power cord is connected to a valid power socket.
B. Whether the fuse tube on the back of the instrument is loose or burnt out.
2. The instrument stopped working after clicking on start button?
A. It may be that the water flow of the instrument is insufficient for it to detect the
flow.
B. Check whether the water tank is short on water.
3. No suction when clicking on the start button?
A. Press the start button first, then press the start button of a single handle.
B. Check whether the twisting knob is at the the minimum suction level, which can be
adjusted.
C. Check whether the time has been set to zero, and adjust the time if it is set to zero.
4. No RF output?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether the treatment area has been cleaned. Grease or oil essential
oil products may cause poor contact between the treatment handle of the instrument
and human body, resulting in no RF output.