5. Troubleshooting & Solutions
1. The instrument cannot be started, but the key light on the back of the instrument
does not work?
A. Make sure the power cord is connected to a valid power socket.
B. Whether the fuse tube on the back of the instrument is loose or burnt out.
2. No RF output of the instrument?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether the treatment area has been cleaned. Grease or oil essential
oil products may cause poor contact between the treatment handle of the instrument
and human body, resulting in no RF output.
3. Weakened RF output?
A. Please check whether the plug connecting the instrument handle and the body is
tightly connected.
B. Please check whether non-conductive grease and other substances are stained on
the treatment handle, which will cause poor contact and weaken the output.
C. Please check whether the product used is the adaptive product specified by the
instrument.
4. No suction or very little suction?
A. Please turn off the instrument and check the filter element specially used for the
instrument, which may need to be replaced.
B. Please check the rubber ring of the handle plug, because the rubber ring may be
worn and cause air leakage.
C. Please check whether the oil filter cup outside the filter is tightened and whether
the rubber ring is worn. It is possible that air leakage at this position may lead to poor
air pressure.
D. If the above methods cannot address the issue, please contact the instrument
dealer for assistance.
5. I can start the instrument, but the screen show’s error message?
A. Please unplug the power plug behind the instrument and wait for about 1 minute
before restarting the instrument.
B. If the above methods cannot be handled, please contact the instrument dealer for
assistance.