unexpected situations that may affect the curative effect.
(7) Please do not use the equipment on the eyes, thyroid, parathyroid, testicles, a
pregnant woman's abdomen, pacemaker, etc.
(8) If you are suffering from an illness, you should use it with caution unless you get
permission from a doctor.
(9) Please turn off the equipment's power switch if no one uses it, and ensure the
main power is off before the exit of related personnel to guarantee the
equipment's safety.
(10) When operating the equipment, start from the lowest energy level and gradually
increase it after the client gets used to it.
(11) You can use the handpieces separately, but you can use them together to
achieve a better effect.
(12) When operating the instrument, it should contact the skin thoroughly to avoid
uneven heating.
(13) You must keep the treated parts moist and avoid operation in dry skin conditions
when operating the instrument.
(14) It's advised to stop taking other weight-loss medicines for 1 to 2 months before
treatment if you are taking them now. The course of treatment shall be extended
if you want to take it immediately.
(15) Using 40K on the head, chest, breast, heart, and back is prohibited.
(16) After the operation, clean the equipment with normal saline to ensure its
cleanliness and hygiene and extend its service life.
4. Common Problems & Solutions
(1) Does the equipment can't be started, and the button lamp on it can't light up?
A. Make sure the power cord is plugged into a good power socket.
B. Check whether the fuse tube in the back is loose or burnt out.
(2) Does the RF doesn't have output, or is it weakened after the equipment is on?
A. Please check whether the handpiece and the plug connecting to the equipment
body are closely linked.
B. The wire inside the handpiece is loose or burnt out.
(3) The equipment doesn't have suction, or the suction is weak?
A. Please check whether the handpiece and the plug connecting to the equipment
body are closely linked.
B. Please check whether the oil filter outside is tightened and whether the rubber
ring is worn out. Air leakage in this place may cause insufficient air pressure.
C. If the solutions mentioned above are not working, please contact the equipment
dealer for assistance.
(4) Is there an error message on the monitor after the equipment is on?
A. Take out the plug from the back, wait about one minute, reinsert the plug, and