(14) Start from the lowest energy level when start and increase the level gradually
after the client got used to it.
(15) The device should contact the skin fully when in use. Otherwise, it may not be
attached to the skin.
(16) After the operation, clean the equipment with normal saline to ensure its
cleanliness and hygiene and extend its service life.
(17) When operating, fasten EMS Pad or Breast Pad onto the client first, turn the
equipment on, and adjust the energy level which should be set from low to high
slowly to make the client feel warm and comfortable.
(18) Product with high oil content must be applied, and the treated parts must be kept
moist.
(19) When using EMS Pad, its whole surface should contact the skin thoroughly and
can't only touch a small area.
(20) The EMS pad(breast pad) can be used separately. Fasten it directly onto parts to
be treated. Set energy parameters before use. The treatment time for a single
part is about 20 minutes. A better effect can be achieved with the combined use
of massage, RF, and ultrasound equipment.
4. Troubleshooting & Solutions
(1) The equipment can't be started, and the buttons in its back are not lighting up?
A. Make sure the power cord is plugged into a suitable power socket.
B. Check whether the fuse in its back is loose or burnt out.
(2) The equipment doesn't have RF output?
A. Please check whether the handpiece and the plug connecting to the equipment
body are closely linked.
B. Please check whether the treated parts have been cleaned. Grease or essential oil
products may cause poor contact between the handpiece and the body resulting
in no output.
(3) The RF output is weakened?
A. Please check whether the handpiece and the plug connecting to the equipment
body are closely linked.
B. Please check whether there is non-conductive grease on the handpiece, which
may cause poor contact weakened output.
C. Please check whether the products used are the adaptive products specified by
the equipment.
(4) The equipment can be started, but there is an error message on the monitor?
A. Take out the plug in its back, wait for about 1 minute, re-plug it, and restart the
equipment.
B. If the solutions mentioned above are not working, please contact the equipment
dealer for assistance.