Practice method: 
A.  To  avoid  proficiency  restrictions,  operators  can  first  test  for  a  small  treatment 
area, then operate on large area. 
B. Operators can test between newspaper or dark papers. Adopt the method, from 
left to right, up to down or choose familiar technique to training. Every affected part 
should  be  treated  1-2  times  (according  to  the  sufferer’s  extent).  Untrained  or 
unqualified  operating  persons  are  not  permitted  to  operate  the  laser  treatment 
systems under any circumstances.   
 

Chapter 8 Notices before operation: 

Appliance preparation: napkin, eye area (for operators and patients), cold injection 
machine or ice bag, sunscreen cream. 
Operation procedure: 
A.  Clean the treatment area 
B.  Cleanly shave off the hair on treatment part (such as: women’s beard and hair 

on the temples) 

C.  Exfoliation (except thin skin) 
D.  After adjustment, have an experiment on normal skin to see whether there is 

change and ask for the patient’s feelings; Build the data for the clients about the 
therapy. 

E.  Choose the skin type. 
F.  Shooting the light two times every time after adjusting the energy. 
G.  Treating 
H.  Cold spray or a cold compress   
I.  Smear the sunscreen cream   
Parts can be treated:   
a.  Parts need to be treated; 
b.  Neck

chest

arm

armpit. 

c.  The serious parts which can be scanned twice. 
Requirement: The temperature of sapphire tip need to reach 5

℃。

 

 

Chapter 9 Notices during the treatment 

a.  Make sure the safety of the sufferer. Sufferer’s eyes need to be closed.   

b.  The  header  should  slightly  touch  skin  when  operating.  Don’t  extrude  the 

skin to avoid burn. 

c.  Don’t see the light directly even you wear the protection glasses. 

d.  Choose  reasonable  technical  parameters.  Operators  should  clearly  notice 

the dangerous of intensive light to skin. Have an experiment after the ear’s 

skin before treating. Observing the phenomenon after 20 minutes. It should 

Summary of Contents for HR-F16

Page 1: ...achine 2 Test machine 3 When you change all water please also use distilled water 4 Must immit DISTILLED WATER Change all water 1time month 5 Please start to adjust parameters and treat patients immed...

Page 2: ...r 5 3 1 Device Structure 5 3 2 Application 5 Chapter 4 Working Conditions 6 Chapter 5 Installation and debugging 6 Chapter 6 Technical Specifications 7 Chapter 7 Operation 7 7 1 Starting 8 7 2 Paramet...

Page 3: ...alcohol ether oxygen and light decomposition such as drug products should be far away peak power laser light radiation 2 1 4 Wrist watch necklace bracelets and other metal jewelry reflecting light are...

Page 4: ...lectric shock 2 4 System Device Safety 2 4 1 Water current device when laser emits there is high density heat in the hand piece So it requires constant cooling water conduction cooling The machine is...

Page 5: ...nal standards connections for power 4 4 The power should be well earthed in accordance with the electrical safety regulations in force 4 5 The machine adaptation voltage range is 200 240V 100 120V If...

Page 6: ...3 The water must be empty during the long term stop using or long distance transportation Change water We advice the machine should change distilled water in one month after using 4 After Loosing the...

Page 7: ...7 Chapter 7 Operation 7 1 The Starting interface 7 2 Click diode laser to choose the Diode Laser treatment interface...

Page 8: ...second Frequency recommendation 4 to 6 hertz Refrigeration Control the skin cooling turn red to cooling turn green means no cooling Click ready to work click standby to finish treatment 7 3 The wavele...

Page 9: ...D After adjustment have an experiment on normal skin to see whether there is change and ask for the patient s feelings Build the data for the clients about the therapy E Choose the skin type F Shootin...

Page 10: ...e face appears light red The lesion feels hotness and sting while the root of hair has a red spot The symptoms can be relieved after cold spray or cold compress At the same time Scabbing and the color...

Page 11: ...ing A The working and treatment principle of the machine B Operation procedure C Common malfunction recognition Don t charge any other costs Please contact our After service department if you had any...