36
Table C.2 – Example Immunity Test Level Compliance
The ME EQUIPMENT is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Class B
Class A
Complies
RF Emissions CISPR 11
Harmonic Emissions IEC
61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
The ME EQUIPMENT is intended for use in the electromagnetic environment specified below.
The customer or the user of the ME EQUIPMENT should assure that it is used in such an environment.
Electrostatic
Discharge (ESD) IEC
61000-4-2
Electrical Fast
Transient/Burst IEC
61000-4-4
Surge
IEC 61000-4-5
±8 kV Contact
±15 kV Air
±2 kV for power
supply lines
±0.5 kV, ±1 kV
line to line
±8 kV Contact
±15 kV Air
±2 kV for power
supply lines
±0.5 kV, ±1 kV
line to line
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance