MVE LN2 Tx Operator'S Manual Download Page 6

6 — Ln2 Tx Instructions

Setup

N2 Gas Connection

The LN2 TX requires a nitrogen gas supply to operate.  The con-

nection for the nitrogen gas supply hose is shown in Figure 3.  

The hose connection is a ½” ODT x 45° SAE flare fitting.  This 

is the standard connection for a nitrogen transfer hose.  It is the 

same connection used to connect a transfer hose to an MVE 

Liquid nitrogen freezer.
 

Figure 3, Nitrogen Gas connection

Connect the nitrogen gas supply hose to the LN2 TX as shown in 

Figure 4.
 

Figure 4, Nitrogen Gas supply hose connected to the LN2 TX

Connect the opposite end of the supply hose to the gas use valve 

of a portable liquid nitrogen cylinder as shown in Figure 5. The 

nitrogen gas supply pressure should be between 22 and 50psi.
Although a liquid cylinder with gas use capability is the preferred 

supply method, a high pressure nitrogen gas supply cylinder can 

be used to supply gas to the LN2 TX.  A suitable pressure regula-

tor and hose adapter fittings will be required.  DO NOT USE 

COMPRESSED AIR to supply the LN2 TX.  The moisture in the 

compressed air will freeze inside the suction head of the LN2 TX, 

blocking the orifices and causing it to stop working.

 

Figure 5, Nitrogen Gas Supply hose connected to Gas Use valve 

on liquid cylinder

Basic Operating Instructions

The suction end of the LN2 TX is shown in Figure 6

Figure 6, Suction end of LN2 TX

Place the suction end of the LN2 TX in to the container that you 

want to remove liquid from as shown in Figure 7, making sure 

that the open end does not sit flat on the bottom of the container.  

If the end sits flat on the inside bottom of the container, the inlet 

will be sealed off and no transfer will take place.
 

Figure 7, Suction end of LN2 TX placed inside supply container

Summary of Contents for LN2 Tx

Page 1: ...fer System OPERATORS MANUAL BioMedical Chart Inc BioMedical Group 2200 Airport Industrial Drive Suite 500 Ball Ground Georgia 30107 This Operators Manual covers use and maintenance of the MVE LN2 TX I...

Page 2: ...g 7 Technical Assistance 7 Table of Figures Figure 1 LN2 TX 3 Figure 2 Typical cryogenic Dewar 3 Figure 3 Nitrogen Gas connection 6 Figure 4 Nitrogen Gas supply hose connected to the LN2 TX 6 Figure 5...

Page 3: ...m area in the transfer system thus minimizing flash loss Once the liquid is entrained in the gas stream it is re pressurized to further minimize flash loss N2 Gas Connection LN2 Outlet Flex Transfer H...

Page 4: ...operated in open and well ventilated areas Persons transferring LN2 should make every effort to protect the eyes and skin from accidental contact with liquid or cold gas issuing from the Dewar Protec...

Page 5: ...ere are additional safety considerations to keep in mind in the handling of O2 These additional considerations are associated with the oxidizer properties of liquid oxygen DO NOT SMOKE NEAR THE TRANSF...

Page 6: ...gas use capability is the preferred supply method a high pressure nitrogen gas supply cylinder can be used to supply gas to the LN2 TX A suitable pressure regula tor and hose adapter fittings will be...

Page 7: ...techservice australia chart ind com The discharge end of the LN2 TX is shown in Figure 8 Figure 8 Discharge end of LN2 TX Place the discharge end of the LN2 TX as shown in Figure 9 The LN2 TX is equip...

Page 8: ...8 Ln2 Tx Instructions...

Page 9: ...PN 15098830 Rev A 9...

Reviews: