Mutable Instruments Streams User Manual Download Page 7

The symptoms of a miscalibration are the following:

The bagraphs display inaccurate values - for example a LED is lit when you monitor a silent channel.
There is some faint signal (“bleed”) even if the level of a channel ought to be null.

If this happens, please follow this procedure:

Set the response curve 

(E)

 of both channels to linear.

Plug 6 patch cables into all inputs, while leaving the other end of the cable unconnected. Hold the METER
button for 1s. The bargraphs show a “C” letter. The bargraphs are now calibrated.

(The bargraphs are still showing a “C” letter)

Connect the output of an oscillator to channel 1 input. Monitor the signal on channel 1 output and adjust
channel 1’s shape potentiometer until you reach a point where the faint signal you can hear (bleed)
disappears. The ideal position is the point at which the bleed just starts disappearing.

(The bargraphs still showing a “C” letter)

Same as step 2, but for channel 2. Once you are done, press any button to write the settings and leave
the calibration mode.

This product is covered by Mutable Instruments’ warranty, for one year following the date of manufacture.
This warranty covers any defect in the manufacturing of this product. This warranty does not cover any
damage or malfunction caused by incorrect use - such as, but not limited to, power cables connected
backwards, excessive voltage levels, or exposure to extreme temperature or moisture levels.

The warranty covers replacement or repair, as decided by Mutable Instruments. Please contact our
customer service ([email protected]) for a return authorization before sending the
module. The cost of sending a module back for servicing is paid for by the customer.

Mutable Instruments encourages modding and hacking, but we will not service modified units or provide
any assistance in the realization of mods.

Mutable Instruments | Streams

7 of 8

Summary of Contents for Streams

Page 1: ...a pulse on the EXCITE input Vactrol emulation the EXCITE input is fed into a digital model of a resistive opto isolator vactrol This generates a slow moody non linear response Envelope and filter fol...

Page 2: ...amplitude response curve This adjusts the response of the VCA from exponential to linear F Metering button Press the button to show the bargraphs Press again to select which signal is to be monitored...

Page 3: ...LED represents a level change of 6dB with the lower two LEDs lit in green and the upper two lit in orange and red When the compressor mode is active when monitoring the output level the bargraph will...

Page 4: ...to further damp the sound Starting from the minimum setting turning the knob clockwise will progressively increase the amount of filter damping with the maximum amount reached when the knob is at noon...

Page 5: ...follower wouldn t show much variation because the waveform is dominated by the drone Band wise processing makes the detection process more sensitive to the drums The spectral centroid is approximated...

Page 6: ...compressor mode you gain complete control over all parameters Channel 1 SHAPE Attack time Channel 2 SHAPE Release time Channel 1 MOD Threshold Channel 2 MOD Ratio Makeup Connect the output of your au...

Page 7: ...starts disappearing The bargraphs still showing a C letter Same as step 2 but for channel 2 Once you are done press any button to write the settings and leave the calibration mode This product is cove...

Page 8: ...Mutable Instruments Streams 8 of 8...

Reviews: