![Murata 4268A Manual Download Page 13](http://html1.mh-extra.com/html/murata/4268a/4268a_manual_1857294013.webp)
- 12 -
”
A
”
field in the above figure indicates both ALC and Hi-PW being set to
“
ON
”
.
An asterisk mark at
“
B
”
field in the figure means ALC being on.
Fig.16 Measure Setup Screen Showing ALC ON
“
Vm
”
in
“
A
”
field in the above figure indicates actual measurement voltage monitored by the meter, which
shows 1V being applied to the sample.
“
B
”
field shows the capacitance of a 10
F product, which is about
9.86
F which is within the specified tolerance.
Fig.17 Measure Setup Screen Showing ALC ON
“
A
”
field in the above figure indicates ALC being set to
“
OFF
”
and Hi-PW being set to
“
ON
”
. No asterisk mark
at
“
B
”
field in the figure means ALC is off.
Fig.18 Measure Setup Screen Showing ALC OFF
B
A
B
A
B
A