3. ADJUSTMENT OF SET POINTS: 

(a) RATE1 [

rat1 

]:

 

 

Press            key for 2 seconds. 

 

You are in SET MODE. 

 

Display 1 will show

 rat1

 

Display 2 will indicate current value of set point. Range is 

0.1 to 9.9°C/min

  

 

You can change current value of this set point by following PROCEDURE. 

 

Increase   the parameter by            key. 

 

Decrease the parameter by            key. 

 

Press             to store it in memory or press             for next step. 

 

If No key is pressed for 5 seconds previous value of parameter is retained and controller goes in control 
mode. 

(b) SET POINT 1 [

Set1]

:

 

 

Press            key for 4 seconds. 

 

You are in SET MODE. 

 

Display 1 will show 

SEt1

 

Display 2 will indicate current value of set point. Range is 

0°C to 1200°C

  

 

You can change current value of this set point by following PROCEDURE. 

 

Increase the parameter by            key. 

 

Decrease the parameter by            key. 

 

Press             to store it in memory or press             for next step. 

 

If No key is pressed for 5 seconds previous value of parameter is retained and controller goes in control 
mode. 

(c) HOLD TIME [

HLd1]

:

 

 

Press            key for 6 seconds. 

 

You are in SET MODE. 

 

Display 1 will show

 HLd1

 

Display 2 will indicate current value of set point. Range is 

1 to 999min.

  

 

You can change current value of this set point by following PROCEDURE. 

 

Increase   the parameter by            key. 

 

Decrease the parameter by            key. 

 

Press             to store it in memory or press             for next step. 

 

If No key is pressed for 5 seconds previous value of parameter is retained and controller goes in control 
mode. 

NOTE:  You  can  do  the  setting  of  the  remaining  parameters  Rate2,  Setpoint2,  Holdtime2  Rate3,  Setpoint3, 

Holdtime3,  Rate4,  Setpoint4,  Holdtime4,  Rate5,  Setpoint5,  Holdtime5,  Rate6,  Setpoint6, 
Holdtime6,  Rate7,  Setpoint7,  Holdtime7,  Rate8,  Setpoint8,  Holdtime8,  &  Rate9,  Setpoint9, 
Holdtime9 just like (a), (b) & (c).  

NOTE: Here, the Control Action is On-Off & Hysteresis provided is fixed which is 2°C. 

NOTE:  

  Initially  as  the  power  is  turned  on  Relay  (R2)  is  turned  ON.  After  completion  of  hold  9  the 

process will be off and r2 goes OFF, when process temperature reaches at ALARM set point 
and lower display will show end. 

 
 
 
 
 
 

WORKING

 

1)

 

As  the  power turns ON upper display shows room temperature/ process temperature and lower 
display shows value current temperature, which changes according to the Rate (°C/min).  

SET

 

 

SET

 

ENT

 

 

SET

 

 

SET

 

ENT

 

 

SET

 

 

SET

 

ENT

 

 