2. Electrical Specifications
Accuracy: ±(% R Digits), one-year calibration period; If the temperature variation of environment
reaches ±5
°
C, the accuracy can be adopted after two hours.
The accuracy can be adopted after two hours when the battery charging is completed.
Ambient temperature: 23
°
C± 5
°
C;
Ambient humidity: .;;; 75%RH;
Temperature coefficient: 0.1x (Accuracy)/
°
C (<18
°
C or >28
°
C)
(1) AC Voltage
Range
Resolution
Accuracy Tolerance:±(% Digits)
60mV
0.001mV 45
""
1 kHz
1 k-10kHz
10k-20kHz
20k-100kHz
±(0.6%+60)
±(1.2%+60)
±(3%+60)
±(4%+60)
600mV 0.01mV
45-1kHz
1 k-10kHz
10k-20kHz
20k-100kHz
±(0.3%+30)
±(1.2%+40)
±(3%+40)
±(4%+40)
6V
0.0001V
45-1kHz
1 k-10kHz
10k-20kHz
20k-100kHz
±(0.3%+30)
±(1.2%+40)
±(3%+40)
±(4%+40)
60V
0.001V
45-1kHz
1 k-10kHz
10k-20kHz
20k-100kHz
±(0.3%+30)
±(1.2%+40)
±(3%+40)
±(4%+40)
600V
0.01V
45-1kHz
1 k-10kHz
10k-20kHz
20k-100kHz
±(0.3%+30)
±(1.2%+40)
±(3%+40)
Only for reference
1000V
0.1V
45-1kHz
1k-5kHz
5k-10kHz
1 0k-1 00kHz
±(0.6%+30)
±(3%+40)
±(6%+40)
Only for reference
• Input impedance: About 10MO
• Overload protection: 1000V
• Display: True virtual value for 10% to 100% of the range.