30
Auto Record
1.
Configure the parameters: Press the front panel
key, press the Auto softkey.
Press the Point softkey to specify the total number of readings to record. The range is
1 to 1000.
Press the Interval softkey to specify the time interval between readings. The range is
15 ms to 9999.999 s.
2.
Record data: Press the Start softkey to start auto record. The
icon will show up
on the top of the display. Press the Stop softkey to stop recording,
the data table
shows the readings taken. press
keys to turn the page.
Note:
Automatic recording function supports switching measurement function.
In auto range, the relay switch may cause jitter, the data at this time is invalid.
It
will last about a few hundred milliseconds, and the data acquired in this period will
be marked as "invalid".
When the dual display is enabled, only the reading of main display function can be
saved.