Smart Digital Storage Oscilloscope 

Models MP720009, MP720010 and MP720011 

 

Summary of Contents for MP720009

Page 1: ...i Smart Digital Storage Oscilloscope Models MP720009 MP720010 and MP720011...

Page 2: ...he Horizontal System 14 Introduction to the Trigger System 15 4 Advanced User Guidebook 16 How to Set the Vertical System 17 Use Mathematical Manipulation Function 18 The Waveform Calculation 19 Using...

Page 3: ...pturing a Single Signal 50 Example 4 Analyze the Details of a Signal 51 Example 5 Application of X Y Function 53 Example 6 Video Signal Trigger 55 7 Troubleshooting 56 8 Technical Specifications 57 Ge...

Page 4: ...ts input or output terminals When the instrument is powered by AC do not measure AC power sources directly otherwise it will cause a short circuit This is because the testing ground and power cord gro...

Page 5: ...could result in damage to this product or other property Terms on the product The following terms may appear on this product Danger Indicates an immediate hazard or injury possibility Warning Indicat...

Page 6: ...er when the AC powered oscilloscope is connected to the AC powered PC through the ports Warning To avoid fire or electrical shock when the oscilloscope input signal connected is more than 42V peak 30V...

Page 7: ...which you can set the different options for the current menu The other buttons are function buttons through which you can enter different function menus or obtain a specific function application dire...

Page 8: ...nts 3 AC power input jack 4 Feet Adjust the tilt angle of the oscilloscope 5 USB Device port It is used to transfer data when external USB equipment connects to the oscilloscope regarded as slave devi...

Page 9: ...sing this button is equal to the Save option in the Save menu The waveform configure or the display screen could be saved according to the chosen type in the Save menu 5 Vertical control area with 3 b...

Page 10: ...for a trigger Scan Capture and display the waveform continuously Stop Data acquisition stopped 4 The two blue dotted lines indicates the vertical position of cursor measurement 5 The T pointer indica...

Page 11: ...he Pulse count PC the Pulse count E the Rise edge count E the Fall edge count AR the Area CA the Cycle area 15 The readings show the record length 16 The frequency of the trigger signal 17 The reading...

Page 12: ...he Function Inspection Make a fast function check to verify the normal operation of the instrument according to the following steps 1 Connect the power cord to a power source Press the button on the b...

Page 13: ...at of the switch in the probe as 10X see How to Set the Probe Attenuation Coefficient on P11 and connect the probe with the CH1 channel If a probe hook tip is used ensure that it keeps in close touch...

Page 14: ...he left menu corresponding to the probe This setting will be valid all the time before it is changed again Caution The default attenuation coefficient of the probe on the instrument is preset to 10X M...

Page 15: ...libration The self calibration application can make the oscilloscope reach the optimum condition rapidly to obtain the most accurate measurement value You can carry out this application program at any...

Page 16: ...ity Vertical offset back to 0 shortcut key Turn the Vertical Position rotary control to change the vertical display position of channel and push the position rotary control to set the vertical display...

Page 17: ...s accordingly 2 Use the Horizontal Position rotary control to adjust the horizontal position of the signal in the waveform window The Horizontal Position rotary control is used to control the triggeri...

Page 18: ...ged 2 Use the Trigger Level rotary control to change the trigger level setting By turning the Trigger Level rotary control the trigger indicator in the screen will move up and down With the movement o...

Page 19: ...Trigger System How to Set the Sampling Display How to Save and Recall Waveform How to Implement the Auxiliary System Function Setting How to Measure Automatically How to Measure with Cursors How to U...

Page 20: ...list Function Menu Setting Description Coupling DC AC Ground Pass both AC and DC components of the input signal Block the DC component of the input signal Disconnect the input signal Inverted ON OFF D...

Page 21: ...f the probe is 1 1 the menu setting of the input channel should be set to1X Take the Channel 1 as an example the attenuation coefficient of the probe is 10 1 the operation steps are shown as follows 1...

Page 22: ...elect Factor1 as CH1 4 In the right menu select Sign as 5 In the right menu select Factor2 as CH2 6 Press Next Page in the right menu Select Vertical div the symbol is in front of div turn the M rotar...

Page 23: ...the math menu in the right 2 In the right menu select Type as FFT 3 In the right menu select Source as CH1 4 In the right menu select Window Select the proper window type in the left menu 5 In the ri...

Page 24: ...etitive signals and measuring frequency components near DC Recommend to use for Transients or bursts the signal level before and after the event are nearly equal Equal amplitude sine waves with freque...

Page 25: ...ce signal To reduce random noise and aliased components in repetitive or single shot events set the oscilloscope acquisition mode to average What is Nyquist frequency The Nyquist frequency is the high...

Page 26: ...atical manipulation the analytic resolution of which changes with the time base 2 Horizontal Scale rotary control it is used to set the horizontal scale factor for setting the main time base or the wi...

Page 27: ...ata while waiting for trigger condition to occur Once it detects a trigger it will acquire enough data continuously to draw the waveform on right of trigger point Trigger control area consists of 1 ro...

Page 28: ...trigger level is 0 00mV Edge menu list Menu Settings Instruction Type Single Set vertical channel trigger type as single trigger Single Edge Set vertical channel single trigger type as edge trigger S...

Page 29: ...SC PAL SECAM Select video modulation Next Page Enter next page Sync Line Field Odd Even Line NO Synchronic trigger in video line Synchronic trigger in video field Synchronic trigger in video odd filed...

Page 30: ...ted The reserved original data is displayed in fade colour and Function Menu Setting Description Acquire Mode Sample Peak Detect Average Normal sampling mode Use to capture maximal and minimal samples...

Page 31: ...ical Position rotary controls of Channel 1 are used to set the horizontal scale and position The Vertical Scale and the Vertical Position rotary controls of Channel 2 are used to set the vertical scal...

Page 32: ...ct which the waveform is saved to or recall from Recall or close the waveform stored in the current object address When the show is ON if the current object address has been used the stored waveform w...

Page 33: ...t the same time The stored waveform called out can not be adjusted In order to save the waveform of CH1 CH2 and Math into the object Wave0 the operation steps should be followed 1 Turn on CH1 CH2 and...

Page 34: ...he screen image can only be stored in USB disk so you should connect a USB disk with the instrument 1 Install the USB disk Insert the USB disk into the 7 USB Host port of Figure 3 1 Front panel If an...

Page 35: ...ation about the USB disk will display on the right side with red mark 1 and 2 Figure 4 2 Disk Management of computer 4 Right click 1 or 2 red mark area choose Format And system will pop up a warning m...

Page 36: ...Download URL http www partitionwizard com free partition manager html Tip There are many tools for the USB disk formatting on the market just take Minitool Partition Wizard for example here 1 Connect...

Page 37: ...t click 1 or 2 red mark area choose Format Figure 4 7 Choose format 5 Set File System FAT32 Cluster size 4096 Click OK Figure 4 8 Format setting 6 Click Apply at the top left of the menu Then click Ye...

Page 38: ...User Manual 35 Figure 4 9 Apply setting 7 Formatting process Figure 4 10 Format process 8 Format the USB disk successfully Figure 4 11 Format successfully...

Page 39: ...of Adjust Menu is shown as the follows Function Menu Setting Description Function Configure Show the configure menu KeyLock Lock all keys Unlock method push Trigger Menu button in trigger control area...

Page 40: ...obe attenuation is good The results contain three circumstances Overflow compensation Good compensation Inadequate compensation According to the checking result users can adjust probe attenuation to t...

Page 41: ...ure type press AddCH2 again to add the selected measure type of CH2 Show OFF CH1 CH2 Hide the window of measures Show all the measures of CH1 on the screen Show all the measures of CH2 on the screen R...

Page 42: ...voltage measurements including Mean PK PK RMS Max Min Vtop Vbase Vamp OverShoot PreShoot Cycle RMS and Cursor RMS Figure 4 13 below shows a pulse with some of the voltage measurement points Figure 4 1...

Page 43: ...provide time parameters auto measurements include Period Frequency Rise Time Fall Time D width D width Duty Duty Delay A B Delay A B and Duty cycle Figure 4 14 shows a pulse with some of the time meas...

Page 44: ...rea of the whole waveform within the screen Cycle Area The area of the first period of waveform on the screen and the unit is voltage second The area above the zero reference namely the vertical offse...

Page 45: ...splayed along the horizontal direction of the screen two blue dotted lines displayed along the vertical direction of the screen Cursor measure window at the left bottom of the screen shows the cursor...

Page 46: ...u Display the Freq and Vamp measurement cursor and menu The horizontal cursors are set as the intersections of the vertical cursors and the waveform Line Type Freq Vamp type Freq Vamp Makes the vertic...

Page 47: ...rsors active Select Line in the right menu as a or b turn the M rotary control to move it 7 In the right cursor menu you can select Window as Main to make the cursors shown in the main window How to U...

Page 48: ...function requires that the frequency of signal should be no lower than 20Hz and the amplitude should be no less than 5mv Otherwise the Autoset function may be invalid Run Stop button Enable or disabl...

Page 49: ...red on 2 Open the Window Device Manager by right clicking on My Computer icon then selecting Manage from the left hand column 3 Click on Device Manager and the list of devices will display on the righ...

Page 50: ...User Manual 47 Figure 5 1 Control with PC through USB port Click on Data Download icon...

Page 51: ...Autoset to make the waveform optimized based on which you can further regulate the vertical and horizontal divisions till the waveform meets your requirement 2 Perform Automatic Measurement The oscil...

Page 52: ...H1 channel with the circuit signal input end and the CH2 channel to the output end Operation Steps 1 Push the Autoset button and the oscilloscope will automatically adjust the waveforms of the two cha...

Page 53: ...type this helps user to make few small adjustments to achieve a proper trigger level and mode Here is how we achieve this The operation steps are as follows 1 Set the probe menu attenuation coefficie...

Page 54: ...tically and record the waveform during the period around the trigger time By turning the Horizontal Position rotary control in the horizontal control area in the panel you can change the horizontal tr...

Page 55: ...e more details about the signal The Average function offered by our Oscilloscope can help you to achieve this Here are the steps for how to enable Average function 1 Push the Acquire button to display...

Page 56: ...ing steps 1 Set the probe menu attenuation coefficient for 10X and that of the switch in the probe for 10X see How to Set the Probe Attenuation Coefficient on P11 2 Connect the probe of channel 1 to t...

Page 57: ...obtained namely q arcsin A B or arcsin C D If the principal axis of the ellipse is in the I and III quadrants the determined phase difference angel should be in the I and IV quadrants that is in the...

Page 58: ...ps 1 Push the Trigger Menu button to display the right menu 2 In the right menu select Type as Single 3 In the right menu select Single as Video 4 In the right menu select Source as CH1 5 In the right...

Page 59: ...Set the Probe Attenuation Coefficient on P11 4 There is a waveform displayed but it is not stable Check whether the Source item in the TRIG MODE menu is in conformity with the signal channel used in...

Page 60: ...Bandwidth MP720009 20 MHz MP720010 50 MHz MP720011 100 MHz Channel 2 channels Acquisition Mode Normal Peak detect Averaging Sample rate real time MP720009 100 MS s MP720010 500 MS s MP720011 1 GS s In...

Page 61: ...accuracy 3 DC accuracy average Delta Volts between any two averages of 16 waveforms acquired with the same scope setup and ambient conditions V 3 reading 0 05 div Waveform inverted ON OFF Measurement...

Page 62: ...acy typical Internal 0 3 div Trigger displacement According to Record length and time base Trigger Holdoff range 100 ns 10 s 50 level setting typical Input signal frequency 50 Hz Edge trigger slope Ri...

Page 63: ...oltage 1 M Frequency Typical Square wave of 1 KHz Power Mains Voltage 100 240 VACRMS 50 60 Hz CAT Power Consumption 15 W Fuse 2 A T class 250 V Environment Temperature Working temperature 0 40 Storage...

Page 64: ...User Manual 61 9 Appendix Appendix A Enclosure The accessories subject to final delivery Standard Accessories Power Cord CD Rom Quick Guide USB Cable Probe Probe Adjust Options Soft Bag...

Page 65: ...tective screen 2 Wipe the instrument in power off status by soft cloth that is lightly moistened Use only mild detergent and fresh water Do not use any abrasive chemicals to avoid the instrument or th...