Range

Max 

measurement 

range

Resolution

Accuracy ±(% r digits)

30V

0V~31V

0.001V

0.02%+2

-10

o

C~18

o

C, +28

o

C~55

o

C temperature coefficient: ±0.005%FS/

o

C

Input resistance: >1MΩ

Range

Max 

measurement 

range

Resolution

Accuracy ±(% r digits)

24mA

0~24mA

0.001mA

0.02%+2

24mA 

Loop

0~24mA

0.001mA

0.02%+2

-10

o

C~8

o

C, +28

o

C~55

o

C temperature coefficient: ±0.005%FS/

o

C

Input resistance: <100Ω

Range

Max 

measurement 

range

Resolution

Accuracy ±(% r digits)

24mA

0~24mA

0.001mA

0.02%+2

24mA 

Simulating 

Transmitter

0~24mA

0.001mA

0.02%+2

-10

o

C~18

o

C, +28

o

C~55

o

C temperature coefficient: ±0.005%FS/

o

C

Max load voltage: 20V, equivalent to the voltage of 20mA current on 1000Ω load.

DC Voltage Measurement

DC Current Measurement

DC Current Output

Max voltage between any terminal and ground or between any two terminals: 30V 

Range:    

 

manual

Operating temperature:   0°C~50°C (32°F~122° F)

Storage temperature:  

-20°C~70°C (-4°F~158° F)

Relative humidity:  

<95% (0°C~30°C), <75% (30°C~40°C), <50% (40°C~50°C)

Operating altitude:  

0-2000m

Battery: 

 

  9Vx1

Dimension:  

 

about 96x193x47mm

Weight:   

 

about 370g (including battery)

24V Power Supply: Accuracy: 10%

SPECIFICATIONS

• 

All specifications are based on a one year calibration period and applied to a 

working temperature range of +18

o

C~+28

o

C unless otherwise specified. 

 

All specifications are obtained after a 30 minute period of operation.

Summary of Contents for MP700005

Page 1: ...1 Loop Calibrator Model No MP700005...

Page 2: ...terminals Children should be supervised to ensure that they do not play with the product Do not use the product for any purpose other than that for which it is designed Do not operate the product aro...

Page 3: ...display testing The LCD screen will display all symbols for 1s during self test Press V button to switch to voltage measurement mode 11 Briefly press the UP or DOWN 25 value buttons to add or subtrac...

Page 4: ...g on the calibrator until the buzzer beeps then adjust the time between 1 30 min with the UP or DOWN 25 value buttons then short press BACKLIGHT button to save the setting if the time is not saved the...

Page 5: ...probe to the positive terminal black to negative terminal Read the data on the LCD screen Current Measurement Short press MEASURE SOURCE and select Measure then press mA button and the LCD displays m...

Page 6: ...e transmitter The steps are as follows Press MEASURE SOURCE button to select the measure function and the LCD will display MEASURE Long press V button and the LCD will display MEASURE LOOP the unit is...

Page 7: ...nal black to COM terminal Connect the red probe to the ammeter positive terminal black to the ammeter negative terminal Select an output digit by using LEFT 0 or RIGHT 100 value buttons and adjust its...

Page 8: ...e LCD will display SOURCE Long press mA button and the LCD will display SOURCE SIM the unit is mA Connect the red test lead to mA terminal black to COM terminal Connect the red probe to positive termi...

Page 9: ...se the output in 25 increments Long press LEFT 0 or RIGHT 100 to switch the output between 0 and 100 range Auto Ramping inc dec the Output The auto ramping function allows you to continuously apply a...

Page 10: ...fficient 0 005 FS o C Max load voltage 20V equivalent to the voltage of 20mA current on 1000 load DC Voltage Measurement DC Current Measurement DC Current Output Max voltage between any terminal and g...

Page 11: ...or the battery cover power the instrument down and remove the test leads from any device or circuit under test and remove the test leads from the input terminals If the low battery symbol illuminates...

Page 12: ...is required Do not dispose of these items with general household waste Separate for the treatment recovery and recycling of the materials used Waste batteries can be returned to any waste battery rec...

Reviews: