OPTICS CLEANING INSTRUCTION 

 

These instructions are provided as a guide only.  MultiCam cannot be held responsible 
for any damage to optics resulting from improper cleaning or handling. Cleaning 
instructions supplied by the optics manufacturer should always be followed.   
 

Read the instructions completely before starting. 
If possible optics should be cleaned in a dust free air-conditioned room. 
Always clean with alcohol all gloves or finger cots before handling any optics. 
Always handle optics by their edges, never touch the optical surfaces. 

Supplies needed: 

1.

 

Latex/non-latex rubber gloves or finger cots 

2.

 

High purity Isopropyl alcohol 

3.

 

Reagent Grade Acetone 

4.

 

Lens tissue 

5.

 

Lint free cloth 

6.

 

Cotton swabs 

7.

 

Glass containers with droppers. 

Cleaning Procedure: 

1.

 

Wash hands with soap to remove all oils, then, put on gloves or finger cots. 

2.

 

Clean gloves/finger cots with lint free cloth saturated with alcohol. 

3.

 

Hold the optic by its edges or secure the optics mount from moving and blow any 
dust off with low-pressure dry nitrogen (2 to 5 PSIG), air from a blow bulb, or 
clean dry air from air compressor.  

CAUTION:

 Do not use air from a shop air 

compressor if water or water vapor is present. 

4.

 

Soak a clean lens tissue with isopropyl alcohol and place on the optic.  If using a 
cotton swab, soak with reagent grade isopropyl alcohol and apply to the face of 
the optic.   

5.

 

With the soaked lens tissue wipe the optic in one direction. If using a cotton swab, 
rotate the swab while moving it across the face with a maximum of 350 degrees 
of rotation.   

6.

 

Repeat steps 3 and 4 three times using a clean lens tissue or cotton ball. 

7.

 

Inspect the optic. If there is any contaminate on it or it appears cloudy repeat steps 
4 through 5. If it is clean, proceed to step 8.  

8.

 

While holding the optic by its edges or securing the optics mount from moving, 
place a clean piece of lens tissue over the optic, then apply a few drops of acetone 
to the lens tissue. 

9.

 

Slowly drag the lens tissue off the surface of the optic in one direction. 

10.

 

Repeat step 8 & 9 three times. 

11.

 

Rotate the optic 90

0

 and repeat steps 8 & 9 three times.  

12.

 

Inspect the optic to make sure that it is clean. 

If cleaning a lens turn the optic over and repeat steps 3 through 11 if required. 

Summary of Contents for 2000L Series

Page 1: ...Collimator beam expander The laser beam is a diverging source The farther away from the source that you get the larger the beam diameter becomes This optical device is used for two purposes 1 to colli...

Page 2: ...in the MultiCam are made from ZnSe and proper cleaning Refer to Optics Cleaning Instructions handling and disposal Handling broken burned lens document should be followed Tools needed for this alignm...

Page 3: ...s active until the CANCEL button is pressed on the keypad Laser Alignment Steps Follow these steps in order and you will have a successful beam alignment As you gain experience some can be combined or...

Page 4: ...the collimator alignment tool The collimator alignment tool is installed using the supplied fasteners stored on the beam bender base attached to the alignment tool to BB1 as shown in the following pic...

Page 5: ...t a distance equal to 2X the total beam travel if possible It will also be necessary to check at 3FT 1M intervals between the near and far field locations It may be necessary to increase the laser pow...

Page 6: ...ext step STEP 9 Install mirror block assembly on BB3 Install the mirror block assembly on BB3 using the three mounting screws Tighten the screws then loosen one full turn to allow for adjustment durin...

Page 7: ...ews then loosen one full turn to allow for adjustment during this procedure Remove the tip retainer by moving the z axis to the up position and unscrew it from the cutting head Insert the tip retainer...

Page 8: ...ide of the tip and slightly depress into the tip opening Set the laser power to 5 power and pulse duration of 0 01 seconds Pulse the laser Remove the tape and note where the laser burn is in reference...

Page 9: ...re dry nitrogen 2 to 5 PSIG air from a blow bulb or clean dry air from air compressor CAUTION Do not use air from a shop air compressor if water or water vapor is present 4 Soak a clean lens tissue wi...

Page 10: ...ds onto food and inserted into the mouth The ZnSe material itself is not toxic but the toxicity studies did not examine the long term effects of ZnSe dust inhalation As always operators should observe...

Reviews: