Multicam 2000L Series Alignment Procedure Download Page 1

Laser Alignment Procedure 

 
This procedure is to be used as a guide for the initial system installation, or as a reference 
for optics replacement, or for re-alignment if necessary.  It is assumed that the gantry is 
square to the x-axis, all support equipment is in place and properly working, and all of the 
necessary safety precautions are strictly followed.  It is strongly suggested that at least 
two people be involved in the alignment procedure. 
 
The MultiCam laser system utilizes a flying optics beam path.  The laser head is 
stationary and the optics guides the laser beam to the work surface.  The components that 
make up the laser beam guidance system are specially designed to give a consistent beam 
diameter at any location across the cut bed when properly aligned.  The system relies on 
the alignment of its beam path components to accurately and consistently cut materials.  
Occasionally, the beam path components must be realigned due to conditions that impact 
the alignment like severe vibration movement around the machine, shifts in the floor, or 
other uncontrollable conditions.  The complete optical system is pre-aligned at the 
MultiCam factory and then the laser head is removed for shipping.  When re-assembling 
the laser optical system, minimal adjustments will have to be made if this procedure is 
strictly followed. 
 

Optics used in beam path 

 

Collimator/beam expander 

The laser beam is a diverging source.  The farther away from the source that you get, the 
larger the beam diameter becomes.  This optical device is used for two purposes.  1) to 
collimate the diverging beam to insure a consistent beam diameter throughout the cutting 
area and 2) to expand the beam so that the beam diameter is roughly 70-80% of the beam 
path optical components clear aperture.  A 2X beam expander is commonly used in the 
laser beam path giving an output beam diameter of 8-10mm.  It consists of two focusing 
lens positioned to give the proper beam diameter at the output.  Loosening the clamp ring 
fastener and sliding one lens in reference to the other is how the adjustment of the 
collimator is done. 

 

Diode pointer/Beam dump 

The diode pointer is used as an aid in doing the laser beam alignment and used as a laser 
beam reference at the work surface when aligning materials or locating the focused laser 
beam.  The beam dump is used as a safety component for the removal of unwanted laser 
radiation.  Alignment of the diode pointer is done using the two sets of adjusting screws.  
One set is used for positioning in the near field and the other for the far field. 

 

Beam Benders 

The beam benders used in the optical path are made up of a beam bender base, mirror 
block assembly, and the mirror holder.  The mirror block assembly is designed to allow 
adjustment in the vertical and horizontal planes as indicated by the arrows on the mirror 
block assembly.  The mirror block assembly mounting screws are spring loaded allowing 

Summary of Contents for 2000L Series

Page 1: ...Collimator beam expander The laser beam is a diverging source The farther away from the source that you get the larger the beam diameter becomes This optical device is used for two purposes 1 to colli...

Page 2: ...in the MultiCam are made from ZnSe and proper cleaning Refer to Optics Cleaning Instructions handling and disposal Handling broken burned lens document should be followed Tools needed for this alignm...

Page 3: ...s active until the CANCEL button is pressed on the keypad Laser Alignment Steps Follow these steps in order and you will have a successful beam alignment As you gain experience some can be combined or...

Page 4: ...the collimator alignment tool The collimator alignment tool is installed using the supplied fasteners stored on the beam bender base attached to the alignment tool to BB1 as shown in the following pic...

Page 5: ...t a distance equal to 2X the total beam travel if possible It will also be necessary to check at 3FT 1M intervals between the near and far field locations It may be necessary to increase the laser pow...

Page 6: ...ext step STEP 9 Install mirror block assembly on BB3 Install the mirror block assembly on BB3 using the three mounting screws Tighten the screws then loosen one full turn to allow for adjustment durin...

Page 7: ...ews then loosen one full turn to allow for adjustment during this procedure Remove the tip retainer by moving the z axis to the up position and unscrew it from the cutting head Insert the tip retainer...

Page 8: ...ide of the tip and slightly depress into the tip opening Set the laser power to 5 power and pulse duration of 0 01 seconds Pulse the laser Remove the tape and note where the laser burn is in reference...

Page 9: ...re dry nitrogen 2 to 5 PSIG air from a blow bulb or clean dry air from air compressor CAUTION Do not use air from a shop air compressor if water or water vapor is present 4 Soak a clean lens tissue wi...

Page 10: ...ds onto food and inserted into the mouth The ZnSe material itself is not toxic but the toxicity studies did not examine the long term effects of ZnSe dust inhalation As always operators should observe...