2) Set range switch to [125V] or [250V]. (By whichever range, AC 0~600V can be measured 

in the same way). 

3) Contact measuring probe (LINE) and clip (EARTH) to the circuit or the live part of 

equipment to be measured. 

4) Read the voltage value after the display got stable.

 

 
 
 
                                                                                                                     
                                                                                                                                ~ 
 
 
 
 
 
     
 
 
 
 
 
 

4. REPAIR SERVICE 

When requesting for repair service, please bring the instrument directly to the dealer where 

you bought. 

When mailing the instrument, always pack it in its original or equivalent packing materials to 

avoid any damage during the transportation and also put together with documents showing 

your name, address, phone number and defect point. 

 

5. WARRANTY 

This instrument is sent out from our factory after the sufficient internal inspections but if you 

find any defect due to the fault in our workmanship or the original parts, please contact the 

dealer where you bought the instrument. 

The warranty period is 12 months from the date of purchase and the instrument shall be 

repaired at free of charge, provided that we judge the cause of defect is obviously resulted 

from our responsibility.

 

Set OFF 

ACV 

!  WARNING 

1) Do not apply the voltage exceeding AC 600V to prevent electric failure and or burnout 

of the instrument. 

2) Do not apply voltage under the condition of measuring switch MEASURE on, as it 

may cause electrical shock and or instrument damage. 

Summary of Contents for MIS-PV2

Page 1: ...ated voltage in PV systems 4 Can read the displayed value in the dark by LED back light of LCD 5 Auto power off function enables min power consumption in case of forgetting switch power off 6 Convenie...

Page 2: ...ter opened the box check the appearance of instrument and confirm the accessories are contained 2 Specifications 2 1 Insulation Resistance Measurement 1 Rating Range Rated Measuring Voltage Max Effect...

Page 3: ...D with bar graph annunciator 13 Over Range OL mark on display Range Rated Measuring Voltage Max Effective Display Value Measuring Range Accuracy 125V 125V 20M First Effective Range 0 02M 10M Second Ef...

Page 4: ...on 2 2 AC Voltage Measurement ACV 1 Measuring Method Average value display 2 Measuring Range AC 0V 600V min resolution 0 1V 3 Accuracy 1 5 rdg 10dgt 2 3 General Specification 1 Operating Temperature...

Page 5: ...e instrument is getting power on during pressing measuring switch By standing up this switch it will be locked as power on continuously Switch Switch Measuring Switch Display Name Plate for Rating Con...

Page 6: ...ved in battery case 3 In case of not using for a long time remove all batteries as it may cause exhaustion and leakage of batteries WARNING 1 Misapplication of battery polarity may cause leakage of ba...

Page 7: ...ghtened and the first measurement finished The first displayed value will be kept until finishing the second display By pressing data hold switch once the display will be hold and will be released by...

Page 8: ...pletion examination at low voltage circuit set the range switch to 500V 3 Confirm the display value is 0M after short circuited LINE probe and EARTH clip and pressed measuring switch on 4 In case that...

Page 9: ...ing your name address phone number and defect point 5 WARRANTY This instrument is sent out from our factory after the sufficient internal inspections but if you find any defect due to the fault in our...

Reviews: