MULTI MEASURING INSTRUMENTS MCL-800IR Instruction Manual Download Page 9

U

NOTE: 

 

The power will become automatically off approx. 10 minutes after the switch 

operation due to auto power off function. 

 

Measurement of line current cannot be done on Ior current mode. 

 

Current measurement ranges of this instrument are auto but it can be changed to 

manual in case that the range cannot be fixed due to input fluctuation. 

Please refer to (5) Change of Current Range for the operation. 

 

(2) Leakage Current Measurement

 

 

1)

 

Measurement at grounding line 

Do the same method as for line current measurement. 

 

2)

 

Measurement for other circuit except for grounding line 

The operations are the same as for line current but clamp CT to 2 wires en bloc in 

case of single phase and 3 wires en bloc in case of 3 phase. 

 

 

(3) Voltage Measurement (500V Range)   

 

1)

 

Press POWER switch once. All display lightens 

and the range becomes Io current measurement. 

 

2)

 

Press Voltage display switch once and the display 

changes to voltage. 

 

3)

 

Set Frequency switch according to the circuit 

to be measured. 

 

4)

 

Insert the test leads into the voltage input   

terminals (Red & White). 

 

5)

 

Connect the test lead clips to the circuit to 

be measured. 

 

6)

 

Read the displayed value. 

 

 

A

E

A

 

WARNING 

 

Do not apply the voltage more than 500V AC to the voltage input terminals 

(~VINPUT).    It may cause breakage 

 

5B

POSSIBLE ELECTRICAL SHOCK 

 

Measuring test leads are consumable articles. Confirm there is no damage at 

insulation coating of the wire before use. When having found unusual point, 

stop the use and repair or replace to the new one. 

U

POSSIBLE BURNING & WOUND 

 

Connect the lead wires correctly & firmly.   

It may cause spark by mis-connection. 

 

Test Lead 

Red 

White 

Summary of Contents for MCL-800IR

Page 1: ...Before you use your new instrument read this instruction manual completely and familiarize yourself thoroughly with all functions and keep this instruction manual carefully to take out whenever you ne...

Page 2: ...al damage can be supposed caused from mis operations A E A WARNING 2BPOSSIBLE ELECTRICAL SHOCK This instrument is for the use of low voltage circuit Do not make measurements of power lines carrying mo...

Page 3: ...min resolution 0 001mA Ior 0 10A 4 range auto min resolution 0 01mA V 0 500V 1 range min resolution 0 1V M Calculated by voltage and current 0 10M min resolution 0 001M Input frequency range 50Hz 60Hz...

Page 4: ...0A 0 001A 1 000 9 999A Ior 10mA 0 001mA 0 9 999mA 1 5 rdg 20dgt 100mA 0 01mA 10 00 99 99mA 1 5 rdg 15dgt 1000mA 0 1mA 100 0 999 9mA 10A 0 001A 1 000 9 999A V 500V 0 1V 0 499 9V 1 0 rdg 10dgt The condu...

Page 5: ...current and clamp method Open Close Lever CT will open by pushing this lever to inside Power Switch POWER By pressing this switch the power becomes on and pressing once again power off Voltage Displa...

Page 6: ...correctly By pressing once the circuit setting is 1 and twice is and three times is Y Frequency Switch Select 50Hz or 60Hz according to the circuit to be measured Data Hold D HOLD By pressing one tim...

Page 7: ...ument B sign will appear on the display when batteries are exhausted and get less than operation voltage Replace to new batteries immediately Do not use the batteries mixed new one and once used and o...

Page 8: ...ot use the instrument leaving the battery cover off If excessive current is applied to the CT the instrument will be heated and damaged Do not apply more than 50Arms to the CT part Do not apply more t...

Page 9: ...case of 3 phase 3 Voltage Measurement 500V Range 1 Press POWER switch once All display lightens and the range becomes Io current measurement 2 Press Voltage display switch once and the display changes...

Page 10: ...the instrument is looking for phase angle of current and voltage Take care for wiring sufficiently Also if frequency setting is not correct the phase angle is changed and the correct measurement canno...

Page 11: ...switch once Check if the input voltage is correct Press current range switch and select Ior circuit mode to be measured Current range switch will be changed as under by pressing one time 5 To measure...

Page 12: ...b In case of Single Phase Three Wires Measuring Mode 1 Ior c In case of Three Phase Three Wires Measuring Mode Ior White Red White Red LOAD LOAD Red Red White White Black Black LOAD...

Page 13: ...large current fluctuation to be measured however the measurement cannot be done In this case fix the current range by manual mode and make measurement Operation of Manual Range 1 Press Power Switch on...

Page 14: ...ng your name address phone number and defect point WARRANTY This instrument is sent out from our factory after the sufficient internal inspections but if you find any defect due to the fault in our wo...

Reviews: