MULTI MEASURING INSTRUMENTS MCL-800D Instruction Manual Download Page 2

Data hold indication 

"D

H" mark on LCD readout 

Data output 

DC 100mV to each range full scale (DC50mV/1000A) 

Sampling time 

2 times/sec. 

Limitation of circuit voltage 

Less than AC 600V 

Withstanding voltage 

AC2000V/1 minute between CT core and outer case 

Operating temperature 

0

  to +40

80%RH (non-condensing) 

Storage temperature 

-10

  to +60

70% RH (non-condensing) 

Power supply 

UM-4 (1.5V)x2   

Power consumption 

Approx. 3mW 

Battery life 

Approx. 350 hours (by alkaline batteries) 

Size 

138(W)x225(H)x37(D)mm 

Weight 

Approx. 500g 

Accessories 
 
 

Batteries (UM-4)……….….2 
Instruction manual……..…1 
Carrying case………...…….1 

Optional accessory 

Cable for recorder output (MR-002) 

 

 
 
 
4. DIMENSIONS AND PANEL FUNCTION     

 

 
 

 

             

     

                                                 

Current transducer (Jaw)

                 

   

                                                                                               

Jaw opening lever 

                                                                                               

Data hold switch 

LCD display 

Range selector switch

                     

     

 

                       

                                     

Power switch 

                                                                 

                         

Wrist strap 

                               

                             

                         

Output terminal for recorder 

     

                       

                                                         

Battery cover 

                                 

                                                         

Filter switch 

                       

                     

               

 

                  ⑧

 

 
 
 
 

 

3. CAUTION 

● 

Before operating this instrument, familiarize yourself with all instructions   
outlined in this manual. 

● 

Always check to make sure that the function switch is set to the proper position. 

● 

When making measurements, use CAUTION as dangerous voltages may be   
present in normally safe areas. 

● 

To avoid electrical shock, use CAUTION when working above 60V DC or 25V AC 
rms.    Such voltages pose a shock hazard.   

● 

Never make measurements with the battery cover OFF. 

● 

Never fail to keep the maximum tolerable input. 

● 

Never operate this instrument if it becomes wet, damp or has any liquid 
condensation build-up on any part of the instrument. 

   

  Never make measurements for uninsulated conductors or bus bars. 

 