Step 5: Set Treatment Pressure and Time
a. Press and hold the
Pressure Knob
while setting the pressure to the desired
pressure. Once setting the pressure, the
Pressure Knob
is locked.
b. Press the
Time
button to set treatment time.
Note:
•
The minimum pressure setting is 20 mmHg.
The maximum pressure setting is 100 mmHg.
•
The minimum treatment time setting is 15 minutes.
The maximum treatment time setting is 60 minutes.
•
Setting the treatment time to
con
continues the treatment until the
Start/Stop
button is pressed or the
Power Switch
is turned off.
Guidelines for Selecting Pressure
Changing the pressure level alters the treatment intensity by increasing the
amount of air inflating each air chamber.
Lower pressure setting applies lighter treatment, and higher-pressure setting
applies more intense treatment (see Table 2 as a guide for pressure selection).
If the treatment is too intense, decrease the pressure.
Table 2 - Pressure Selection for Arm Sleeve, Full Leg Boot and Calf Boot
Treatment
Intensity
Pressure Setting
(mmHg)
Light
20
–
35
Medium
40
–
65
Intense
70
–
100
Step 6: Select Hold Time
At the end of each inflation cycle, the console holds the air pressure for a few
seconds before releasing the air, pausing, and then continuing to the next cycle
(the
Hold Time
).
You can set the duration of the
Hold Time
for your comfort and to alter the
treatment pace (see Table 3).
Table 3 - Setting Hold Time
Hold Time
Action
Alarm Beeps
Short
Press and hold the
Time
button for 3 seconds.
One beep
Medium
Release and press and hold the
Time
button for 3
more seconds.
Two beeps
Long
Release and press and hold the
Time
button for 3
more seconds.
Three beeps