background image

 

L98107R0-G

 

 

Revive M2 Model 760R 

User Manual 

 

  

Quench Drive 1 
Prairie Du Sac, WI 53578 
+1-608-643-8530, +1-800-356-9522 

www.muellersportsmed.com

 

Summary of Contents for 760R

Page 1: ...L98107R0 G Revive M2 Model 760R User Manual Quench Drive 1 Prairie Du Sac WI 53578 1 608 643 8530 1 800 356 9522 www muellersportsmed com...

Page 2: ...and Dimensions 3 3 Labels 4 4 Indications for Use 6 5 About Revive M2 6 6 Components of the Revive M2 System 7 7 How to Operate the Revive M2 System 10 8 Maintenance and Storage 18 9 Troubleshooting...

Page 3: ...use Safety Warnings and Precautions Warning Explosion Hazard Do not use the Revive M2 system in the presence of flammable gases including flammable anesthetics Electric Shock Hazard Do not allow liqu...

Page 4: ...lint and dirt Keep away from sources of heat or moisture Do not inflate the garment without wearing it over the intended body area Doing so can damage the garment To prevent any damage to equipment u...

Page 5: ...gainst ingress of water IP21 Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Equipment not suitable for use in the presence o...

Page 6: ...no provision for fault protection On console base only in the EU Read instructions before use On garment label Level of protection type BF applied part On console base Double insulation On DC power ad...

Page 7: ...y from sunlight On package Transportation storage humidity limitation On package Transportation storage atmospheric pressure limitation On package Transportation storage temperature limitation On pack...

Page 8: ...ly increases circulation in the areas to which the garment is applied and temporarily relieves muscle aches and pain caused by fatigue or overexertion The Revive M2 Treatment Cycle The Revive M2 Conso...

Page 9: ...h to conserve battery power when not charging or not using the console The following table illustrates the battery power capacity displayed by the battery status indicators Battery Status Displayed Ic...

Page 10: ...eight 8 hours when the battery is fully charged Battery may be charged during normal operation of the console Battery charging time is up to 8 hours Expected service life of battery is two 2 years Re...

Page 11: ...l L98107R0 G 9 Treatment Garments See Figure 6 Figure 7 and Figure 8 A variety of garments are available Leg garments are available in several sizes REVIVE is adjustable Figure 6 Revive Full Leg Boots...

Page 12: ...of set time 15 30 45 60 min or con Start Stop button Used to start or stop treatment Battery Status Indicators displays battery power level Status Indicator White power is On Treatment has not starte...

Page 13: ...surface b Position the console so that you can easily reach the controls during treatment c Select the plug suitable for your country and insert it onto the pins on the DC power adapter backside see f...

Page 14: ...er the intended body area on empty or with open zippers Doing so can damage the garment Do not apply excessive force to the garment straps Do not use the straps for any purpose other than intended by...

Page 15: ...before rising from the treatment position All Garments Make sure the garment does not fit too tightly by inserting three fingers into the garment when it is zipped closed before inflation If the garm...

Page 16: ...nto the air socket on the console see Figure 11 b Make sure the air hoses are not bent kinked or pinched Figure 11 Connecting the Garment Hose Bundle to the console Note When only one Arm Sleeve Full...

Page 17: ...ch air chamber Lower pressure setting applies lighter treatment and higher pressure setting applies more intense treatment see Table 2 as a guide for pressure selection If the treatment is too intense...

Page 18: ...may take several inflation cycles to fill completely as the garment must be filled from a completely empty state i Wait for a few cycles to become accustomed to the treatment sensations Note that the...

Page 19: ...ble 4 Table 4 Stop Restart Pause or Resume the Treatment Treatment Operation Action Stop Press the Start Stop button Restart from the beginning of the selected time Press the Start Stop button again P...

Page 20: ...nlets of the garment d Wipe dry using a soft towel only e When dry the external and internal surfaces of the garment may be wiped with 70 alcohol or with appropriate alcohol based fabric safe disinfec...

Page 21: ...70 C Humidity range of 10 rH to 93 rH non condensing Atmospheric pressure range of 190hPa to 1060hPa Allow the console to reach reasonable room temperature 50 F to 86 F 10 C to 30 C before operating W...

Page 22: ...To replace the battery a Open the battery compartment cover located at the bottom side of the console under the carrying handle see figures below b Disconnect the battery and remove it from the batte...

Page 23: ...s not receive air Examine its hoses for kinks twists and folds The console stops working the Status Indicator turns On in orange and the alarm beeps Hose is not connected properly to garment or consol...

Page 24: ...ts for a period of twelve 12 months from date of purchase This warranty does not include or cover malfunctions caused by unreasonable use noncompliance with user and maintenance instructions or damage...

Page 25: ...tion distances between portable and mobile RF communication equipment and the Revive M2 system Wireless communications equipment such as wireless home network devices mobile phones cordless telephones...

Page 26: ...Model 760R is intended for use in the electromagnetic environment specified below The customer or the user of the Model 760R should assure that it is used in such an environment Immunity Test IEC 606...

Page 27: ...tically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which...

Reviews: