Maintenance
116
Landmark™ Test System Operation
Routine Maintenance Overview Checklist
Calendar Time using 8 hour
Running Time rate per day
Daily Weekly Biweekly
Annually
Running Time-Hours
8
40
80
500
1000 1,500 2,000
Check that Actuator Platen Area is
Clean
X
1
Monitor Filter Indicators
X
Check Hoses/Cables/Connectors
X
Check Crosshead/Lifts/Supports
X
Check Actuator to be Dry
X
Check Hydraulic Service Manifold
X
Check Lift Seal Condition to be Dry
X
Check Lock Seal Condition to be Dry
X
Bionix Lubricate Axial/Torsional
Spline (75-100 hrs)
X
Check Actuator
X
Label Inspection
X
Crosshead/Load Frame
Check Crosshead/Lifts/Supports
MTS
2
MTS MTS MTS
MTS
Lift Seal Condition is Dry
MTS
MTS MTS MTS
MTS
Lock Seal Condition is Dry
MTS
MTS MTS MTS
MTS
Crosshead Columns are Clean
MTS
MTS MTS MTS
MTS
Column Abrasions are Acceptable
MTS
MTS MTS MTS
MTS
Crosshead Speed is Appropriate
MTS
MTS MTS MTS
MTS
Recommended service to be performed at each running time interval noted
1
Symbol denotes services performed by equipment operators. Most of these procedures involve visual
checks that should not interfere with test system operation. These checks are also completed by
trained field service engineers on each Routine Maintenance visit.
2
Symbol denotes service performed by trained field service engineers as part of an MTS Routine
Maintenance plan. Some of these procedures require special service tools and/or specific service
training to complete.