Datalogger Setup
Note:
Any stored data will be permanently erased when
Setup
is finished. To enable you to
save data before it is lost, click on
ʺCancelʺ
and download the data.
Always ensure that the remaining charge in the battery is sufficient to last the complete
duration of your logging exercise. If in doubt, we recommend that you always install a
fresh battery before logging critical data.
Click on the icon on the menu bar. The
Setup
window will appear as shown
below; descriptions for each field in the
Setup
window are listed below the
illustration.
The
Samprate Setup
field instructs the datalogger to log readings at a specific rate.
You can input specific data at the left
Edit
box and select the time unit at the right
Combo
box.
The
LED Flash Cycle Setup
field can be set by the user (select
ʺNo Lightʺ
, LED
will not flash and battery life will last longer).
The
Manual
and
Automatic
options allow the user to start datalogging immediately
when the
Setup
window is exited (automatic) or at a later time (manual).
The
Alarm Setup
field allows the user to set
HIGH
and
LOW
temperature and
relative humidity limits and disables the flashing data logger’s LEDs thereby
conserving the battery energy.
The
Circulating Record
and
No Circulating
options instruct the datalogger to
continue logging data to cover the earlier data (Circulating Record) or stop logging
(No Circulating) when the datalogger is full.
Click on
ʺSetupʺ
to save changes. Click on
ʺDefaultʺ
to set the logger to factory
default condition. Click on
ʺCancelʺ
to abort the setup.
5