background image

5

2.Select the lighting source you want on the light illuminator. If you need UV lighting, select UV (as
shown in Figure 09).

Figure 09

3.According to the diagram on the LED module, pull the pull rod to switch the fluorescent excitation
block to the band matching the light source. If the light source selects UV lighting, the excitation block
also switches to UV.

4.Adjust the brightness knob of the light source control box to adjust the illumination brightness of the
light source according to the fluorescence intensity of the sample.

5.When it is necessary to use microscope bright field lighting, adjust the light source of the light source

control box to the minimum, turn off the light source; then pull the rod of the excitation block of the

fluorescent module to the UV / O position, and then operate the light field observation normally.

Attentions:

1. The brightness of the LED in the illuminator can be adjusted freely according to the difficulty of

excitation of the observed sample; generally, lower than the full load current is beneficial to prolong the

service life of the LED lamp bead; if allowed, it is recommended to adjust the brightness below full load

for use.

2. During fluorescent observation, the eyepiece should not directly face the light or natural light; adjust

the position of the condenser to the lowest, and reduce the aperture of the condenser. If stray light cannot

be eliminated, use a non reflective black plate to block the upper end of the condenser to prevent the

reflection from interfering with the observation.

3. In the fluorescence observation, avoid the sample fluorescence quenching caused by too long

illumination. Adjust the brightness to the minimum or turn off the power switch directly in the

observation gap, and then adjust the brightness when observing.

4. Adjust the excitation block to UV / 0, and then turn off the power supply of the fluorescent module;

Summary of Contents for MI-BGU-LED-TS100

Page 1: ...MI BGU LED TS100 LED Fluorescence Illuminator User Manual ...

Page 2: ...MI BGU LED TS100 LED fluorescence illuminator ...

Page 3: ...er to ensure your safety prolong the life of this unit and maintain it properly please read this manual carefully before operating this unit Warning Do not use or place the instrument in the place with high temperature humidity or dust for a long time Suitable working temperature 5 to 35 Suitable relative humidity 20 to 80 25 Note do not immerse the instrument in water or solvent Note do not place...

Page 4: ...Content Ⅰ Introduction 1 Ⅱ Main specification 2 Ⅲ Components and functions 2 Ⅳ Installation 2 4 Ⅴ Operation 4 5 ...

Page 5: ...rvation Ⅱ Main specification 1 Specification sheet Excitation group LED wavelength Excitation wavelength Dichroic mirror Emmision filter B 470 475nm 450 490nm 505nm 515nmLP G 530 535nm 510 550nm 565nm 575nmLP UV 365nm 330 380nm 400nm 420nmLP 2 Optional specification Excitation group LED wavelength Excitation wavelength Dichroic mirror Emmision filter UV BP 365nm 340 390nm 400nm 450 65nm B BP 470 4...

Page 6: ...roscope 4 Lenses tube connecting LED module with light source and for lighting 5 Power control box exchange current and voltage connecting power control light ON OFF and brightness Ⅳ Installation 1 Loosen the fixing screws of waterproof baffle on the left and right sides of Nikon TS100 microscope host as shown in Figure 01 with an internal hexagonal screw driver and remove the plastic baffle plate...

Page 7: ...nstallation area of the fixed metal plate from the side as shown in Fig 03 There is a screw hole bit on the edge of both ends of the fixed metal plate Use the fixed screw to connect the metal plate and the module body and lock and fix it as shown in Fig 04 Figure 03 Figure 04 4 Insert in the fixing lenses tube from the bottom rear end of the microscope host connect the main body of the fluorescent...

Page 8: ...he two hexagon socket fixing screws on the lens fixing cylinder as shown in Fig 05 Figure 06 7 Connect the light source control box with the light source illuminator of the module and the installation is completed as shown in Figure 07 圖 07 Figure 07 Ⅴ Installation 1 Turn on the power and turn on the power control box switch Figure 08 Figure 08 ...

Page 9: ...ntions 1 The brightness of the LED in the illuminator can be adjusted freely according to the difficulty of excitation of the observed sample generally lower than the full load current is beneficial to prolong the service life of the LED lamp bead if allowed it is recommended to adjust the brightness below full load for use 2 During fluorescent observation the eyepiece should not directly face the...

Page 10: ...to the smoothness of the module position otherwise the uniformity of the light may become worse This power supply must strictly use the special adapter of 12V 2A and the customer who uses other adapters shall be responsible for the equipment damage ...

Reviews: