Appendix
51
Fuses
Electrical fuses are only accessible by MSD service engineers and are not user-serviceable. The information below is provided for
reference.
Main Fuses F1 and F2:
SECTOR Imager 6000 (model 1201) and SECTOR Imager 2400/2400A (models 1250/1251)
T4.0 A, 250 V, 100–240 V~
SECTOR Imager 6000 (model 1200)
T4.0 A, 250 V, 100/120 V~
T2.5 A, 250 V, 220/240 V~
Internal Fuse F3:
SECTOR Imager 6000 (model 1200 only)
T0.8 A, 250 V
8.2.5
SECTOR Imager Physical Dimensions
Size
26.8 in x 30.5 in x 33.9 in (68.1 cm x 77.5 cm x 86.1 cm)
Weight for SECTOR Imager 2400/2400A:
140 lbs (64 kg)
Weight for SECTOR Imager 6000:
150 lbs (68 kg)
8.2.6
Plate Standard Read Volume (per Well)
96-well MULTI-ARRAY Plates
150
µ
L
384-well, MULTI-ARRAY Plates
40
µ
L
1536-well, MULTI-ARRAY Plates
4–6
µ
L
8.2.7
Custom Bar Code Compatibility*
Scientific and Technical Support
for information on proper label, font size, and location of custom bar code on plate.
Recognition of additional codes is possible but has not been tested.
8.2.8
Plate Specifications
MSD MULTI-ARRAY and MULTI-SPOT plates are designed to comply with the mechanical dimensions of the proposed standards of
the Society of Biomolecular Screening. Contact MSD
Scientific and Technical Support
for further details.
The MULTI-ARRAY and MULTI-SPOT plate bar code labels conform to the ANSI/AIM BC1—1995: Uniform Symbology
Specification—Code 39 format.
8.3
Regulatory and Safety Certifications
The SECTOR Imager 6000 and the SECTOR Imager 2400/2400A have been tested to comply with applicable regulatory standards: TUV
(Canada, US), TUV-GS, CE approved, CB Certification.
Regarding EN61326 Electrical Equipment for Measurement, Control, and Laboratory Use—EMC Requirements. (EMCD) Annex B:
•
SECTOR Imager 6000 (model 1200 only) and SECTOR Imager 2400/2400A: These units are designed for operation in a
controlled electromagnetic environment. Transmitters of RF energy such as mobile (cellular) telephones should not be used
in close proximity.
Summary of Contents for SECTOR Image 2400
Page 1: ...SECTOR Imager Models2400 6000 INSTRUMENT MANUAL www mesoscale com ...
Page 6: ...6 1 Introduction ...
Page 9: ...9 2 ImportantInformation ...
Page 15: ...15 3 SystemDescription ...
Page 28: ...28 4 Installation ...
Page 30: ...30 5 QuickStart ...
Page 36: ...Quick Start 36 Figure 5 8 SECTOR Imager window Run Options dialog box ...
Page 38: ...38 6 UsingtheSECTORImager ...
Page 42: ...42 7 Maintenance ...
Page 45: ...45 8 Appendix ...
Page 54: ...54 9 TechnicalSupport ...
Page 56: ...Technical Support 9 2 ...