
7. The sensor is a sealed unit which contains an electrolyte.
Should the sensor develop leakage, it should be
immediately removed from the sensing head and properly
discarded in accordance with local, state and federal
laws. To obtain a replacement sensor, contact MSA at the
address given in Chapter 5, "Obtaining Replacement
Parts." Caution must be exercised so that the electrolyte
does not contact skin eyes, clothing or circuitry. Get
immediate medical attention if electrolyte contacts skin or
eyes.
8. Perform periodic leak check on the sample line and all flow
system components and fittings of the Toxgard II monitor.
9. Use only genuine MSA replacement parts when performing
any maintenance procedures provided in this manual.
Failure to do so may seriously impair instrument
performance. Repair or alteration of the Toxgard II unit,
beyond the scope of these maintenance instructions, or
by anyone other than an authorized MSA serviceman,
could cause the product to fail to perform as designed,
and persons who rely on this product for their safety could
sustain severe personal injury or death.
10. Properly vent the exhaust of the Toxgard II monitor to a
safe area. Improper venting of the exhaust can cause
personal injury or death.
11. Do not install additional filters into the sample inlet fittings.
Adding the filters will impair instrument performance.
12. Ensure the gas sample and Toxgard II unit are at the same
temperature; otherwise, condensation will form, possibly
clogging or filling the sample line.
13. Do not mix units with different area classifications. All units
used in a classified area must have the proper area
classification. Otherwise, an explosion hazard exists,
resulting in possible explosion, injury or death.
FAILURE TO COMPLY WITH THE ABOVE WARNINGS CAN RESULT
IN SERIOUS PERSONAL INJURY OR DEATH.