(2)
Inspect the cartridges to be sure that they are not damaged and the gaskets are
present.
(3)
To ensure proper attachment and prevent cross threading, carefully attach both
cartridges by using the following procedure:
a) Place the unit on a flat surface with the cartridge receptacles facing up
b) Ensure the optional cartridge receptacle plugs are removed
c) Set the cartridge in place on top of the cartridge receptacle
d) Turn the cartridge counter-clockwise until it drops into position
e) Turn the filter clockwise until the cartridge is snug
f) Hand-tighten only
WARNING!
DO NOT cross-thread the cartridges.
Failure to follow this warning can result in serious personal injury or death.
3.6
Optional Accessory
3.6.1 Attaching the Spark Covers (Recommended for use in spark environments)
NOTE:
The spark cover should only be used with the Extended Life Battery (P/N 10076109)
(1)
Remove the cartridges and spark covers from the packaging.
(2)
Attach the cartridges to the PAPR.
(3)
Once the cartridge is attached, align the feet of the spark cover with the inlet hole of
the cartridge.